Product Main

Specifications

100mA photography x ray machine YSX0101
USE

General radiography, Filter-ray-filter radiograph

Unit Configuration 
Control, HT generator unit, Ht cable, X-ray tube unit, Radiographic table-X-ray source column, Collimator 

Function
This photography x ray can used for general radiography, Bucky radiography of body’s head, chest, foot and hand, bones in lying

Feature
1. meter-indicating
2. single-table single tube
3. single-phase, full wave rectification
4. stationary anode x-ray tube assembly
5. X-ray tube filament s voltage regulator, space charge compensator
6. Radiograph plain table wimple structure and easy assembling

Min Technical Indexes

Item
Contents
Index
Demand of power supply
Capacity
Two-phase 30KVA
Voltage, Frequency
380V/220V   ±10%50HZ ±0.5Hz
Intemal resistance
2Ω(380V)  0.6  Ω(220V)
power supply
Adjustable range
380±10% at220V being adjustable continuously
Radiograph
Voltage
50-90KV
Current
Small focus: 15mA,Big focus:30mA 60mA 100mA
Time
0.04~6.35s 23grades together grading a according to R10 coefficient
High-voltage generator
Capacity
7KVA
Max DC output voltage
90KV
Max DC output current
100mA
X-Ray tube
Model
0XD4-2.9/100
Focus
Small focus 1.8×1.8mm  big focus4.3×4.3mm
Radiographic table
Length×width×height of bed surface
2000mm×665mm×700mm
Moving range of bed surface
640mm in Length,200mm in breadth
Ray-filter of table
Travel in length ≥50mm
Grid density N=28
Convergence distance f. =100cm
Grid ratio r=8
Column of X-ray tube unit
Moving Longitudinally along
1000mm or 1800mm
Radiographic table
Distance to friend as moving up and down (along column)
650~1800mm
Rotary around the center of cross arm
±180°
Rotary around the shaft line of X-ray annular tubes
  -10~60~+120°
Max size of cassette
356mm×432mm(14″×17″)

Note: We will not make another notice if the contour of product and technical parameters will be changed