125mA China Radiograhpy X Ray Machines YSX0104
USE
General radiography, Filter-ray device radiography
Unit Configuration
Control, HT generator unit, Ht cable, X-ray tube unit, Diagnostic table, Collimator
Main Characters
1. Single table and single tube
2. Design according to the electric mechanism of middle-size machine. The function of this photography x ray is reasonable.
3. High-tension system is of split structure, with high-voltage generator of full-wave rectification and stable operation
4. Supply voltage (V), photograph kilovolt (kv) and stepless adjust able
5. Equipped with the manostat for filament of X-ray tube and the spacecharge compensator as to ensure the stable and accurate output
6. Equipped with the capacity limitation and portection circuit and KV mA S interlock limitation and protection
7. The timer circuit grade according to R10 priority coefficient and an exact time control
8. The radiograph flat table has a vibrating ray-filter
9. The pillar hast has only the bottom track only, with easy installation
Min Technical Indexes
Supply power
|
Voltage
|
single phase 220V/380V±10%
|
||
Frequency
|
50Hz±0.5Hz
|
|||
Capacity
|
10KVA
|
|||
Intemal resistance
|
380v: 2Ω, 220V: 0.6Ω (2Ω at 80v; 06Ω at 220V)
|
|||
Max DC output voltage
|
90KV
|
|||
Max DC output current
|
125mA
|
|||
Radiograph
|
Voltage
|
50-90KV
|
||
current
|
Small focus 15mA, big focus 30mA 60mA 100mA 125mA
|
|||
spot film
|
100mA(fixed)
|
|||
time
|
0.04-6.3s, 24grades together
|
|||
High-voltage generator
|
Capacity
|
7KVA
|
||
X-Ray tube
|
Model
|
XD4-2.9/100
|
||
Focus
|
Small focus 1.8mm big focus4.3mm
|
|||
Radiograph table
|
Bed face(L×W×H)
|
1960×660mm×700mm
|
||
Bucky
|
Grid density N28
|
|||
Convergence distance fo 100cm
|
||||
Grid ratio
|
||||
Max size of cassette
|
356mm×432mm(14″×17″)
|
|||
Pollar of X-ray source unit
|
Moving in length (along The bottom table )
|
﹥1350mm
|
||
moving un breath
|
﹥360mm
|
|||
moving up and down
|
720~1800mm
|
|||
MAX VOLUME OF RADIOGRAPH
|
||||
Focus of X-Ray tube
|
Radiograph mA
|
Max Radiograph KV
|
Max exposure time allow
|
|
Small focus
|
15mA
|
90KV
|
6.3S
|
|
Large focus
|
30mA
|
90KV
|
6.3S
|
|
60mA
|
90KV
|
6.3S
|
||
100mA
|
85KV
|
2.5S
|
||
125mA
|
82 KV
|
1.0S
|
Note: We will not make another notice if the contour of product and technical parameters will be changed.