Product Main

Specifications

50mA Conventional Diagnostic X-ray Machine YSX0201
Introduction: It can apply for fluoroscopy and radiograph in the hospital and clinic. It also can be used for gastrointestinal examination. It has real-time radiograph device, Electric-line filter, limit beam positioning device, surface-moving electric diagnostic table and automatic overload protection circuit. An ideal multi-functional X-ray diagnostic machine.

GENERAL
It is suitable for both fluoroscopy and radiography in hospitals and health centers. It can also be used in conventional stomach examinations. The unit is provided with a spot-film device, a motor-driven Buck, a collimator and a motor-driven diagnostic table with the tabletop movable.  It is an ideal x-ray diagnostic unit with multi-function
 
MAIN TECHNICAL PARAMETER
X-ray generator
Stationary anode, double focus, and bridge silicon rectification, The X-ray generator is an oil-immersed self-cooled unit.
Max rated capability
Fluoroscopy
75kVp,3mA
Radiography
Current
50mA
30mA
Voltage
90kVp
90kVp
Time
2s
6.3s
Power supply
Voltage
180~240V
Frequency
50Hz
Power
not lower than 5kVA
Resistance
not larger than 1Ω
Voltage adjustment range of x-ray tube
45~90kVp  (continuous)
Range of the timer
0.08~6.3 s
Fluorescent screen size
356mm×356mm
Distance between the focus to the screen
Max 830mm
min 630mm 
Tube head carriage movement range
Vertical: 1100mm
Horizontal: 260mm
Specification of X-ray tube
XD6-3.5/100 Stationary anode, double focus, 2.6mm×2.6mm/1.0mm×1.0mm