Product Main

Specifications

  • YL-2206QK UV weathering chamber
    Purpose:
    UV weathering chamber adopts fluoresvent UV lamps as light sources to
    perform accelerated weathering test on materials by means of simulating
    ultraviolet radiation and condensation of nature’s sunshine,so as to know
    about the weather ability of materials
    This machine is apply to nonmetal material, organic material (like paint
    coating. rubber, plastic and products made in these materials) .
    Standards: GB/T14522-93,GB/T16422.3-1997,GB/T16585-96 etc.
    Technical Parameters: 
    1, Temp. Range: RT+10℃~70℃
    2, Humidity Range: ≥95%R.H
    3, Temp. Accuracy: ≤±0.5℃  
    4, Temp. Uniformity: ≤±2℃
    5, Humidity Accuracy: ≤±2%
    6, Humidity Uniformity: ≤±2%
    7, Blackboard Temp.: 40℃~65℃
    8, Distance Between Lamps: 35mm
    9, Distance Between Lamp and Specimen: 50mm
    10, Lamp Power: 40W
    11, UV Wavelength: 290nm~400nm(Specify when ordering)  
    12, Simulation Condensation: Condensation system time is adjustable
    13, UV exposure, simulation condensation time range: 0~99 hours
    14, Support Specimen: 150×75(mm)
    15, Specimen Quantity: Approx 40pcs
    16, Testing Time: 0~9999H, adjustable 
    17, Irradiance Range: ≤50w/m².
    18, Working Chamber Dimension: D×W×H 450×1170×500mm 

    Light Source:
    1, Using 8pcs UV fluorescent lamps with 40W power rating as light source. UV fluorescent tubes are distributed on both sides of the machine, each side has 4pcs. It has UVA-340 and UVB-313 optional for customers’ configuration. (Option one type)
    2, UVA-340 lamp luminescence spectra energy is mainly concentrated in the 340nmof wavelength.
    3, UVB-313 lamp luminescence spectra is mainly concentrated near the 313nm of wavelength.
    4, Due to fluorescent light energy output will gradually decay over time, in order to reduce the test effect because of light energy attenuation, when in every 1/4 of the fluorescent lifetime of eight lamps in the chamber, a new lamp will replace an old lamp, thus UV light source is always formed by new lamp and old lamp, to get a constant output of light energy.
    5, the effective useful life of the lamp is about 500 hours (imported one is about 1600~1800 hours).
    Standard: GB/T14522-93 GB/T16422.3-1997 GB/T16585-96; and other current UV aging test standards