Product Main

Specifications

Description
DTJS-6009 is designed to measure the Capacitance and Dissipation Factor (tgδ ) of dielectric and electric equipments. It can test grounded equipments, ungrounded equipments, and completely-sealed CVT (Capacitive Voltage Transformer).
The instrument generates testing High-voltage by itself, and the frequency is 45Hz and 55Hz to avoid electromagnetic interference from power frequency, so as to assure the measuring accuracy.

Features
Non-power-frequency Testing voltage, Interference suppression circuit, to assure measuring Accuracy.
Testing voltage is Program-controlled.
Auto balance measuring bridge and calculate dissipation factor to speed and simplify test process.
Large LCD screen.
Menu operation.
Built-in Printer.
Auto identify earthing to ensure safety.
Multi-protection of short-circuit, over-heat, over-voltage, over-current, voltage-fluctuation.

Specifications (English version available)
Testing High-Voltage:
Power: 1kVA
Range: 1~10kV, step 0.1kV
Frequency: 45Hz, 55Hz
Current: max 140mA
Accuracy: 5%
Driving Voltage for CVT:
Range: 0~30V
Frequency: 45Hz, 55Hz
Current: max 15A
Measuring range: tgδ : no limit, Cx: 15pF ~ 100nF
Resolution: tgδ : 0.001%, Cx: 0.1pF
Measuring Accuracy:
Δ tgδ : Reading× 1.0%+0.04%, Δ Cx: Reading× 1.0%+1.0pF
Power supply: AC220V± 10%, 50Hz
Operating conditions: -5~40° C, RH< 80%
Dimension: 47× 32× 39cm (without test leads)
Weight: appr. 30kg (without test leads)
Test leads length: 8m, 8m, 5m