High-speed centrifuge is widely used in the department of research and education of biology, chemical, medicine, and so on. It is a conventional instrument for centrifugation in Lab. The max speed is 8000r/min. It is fit for judging radio-immunity, isolating cell and big particles.
 
Technical Data
A centrifuge is a piece of equipment, generally driven by a motor, that puts an object in rotation around a fixed axis, applying a force perpendicular to the axis. The centrifuge works using the sedimentation principle, where the centripetal acceleration is used to evenly distribute substances (usually present in a solution for small scale applications) of greater and lesser density.
 
Model: YXJ-2
Power Supply: AC220V 50Hz
Maximum Speed: 16000rpm
Max RCF: 17000g
Noise: ≤75dB
Time Range: 0—60min
Capacity: 0.5ml×12, 1.5ml×12
 
Before using should know
1.         Please read the operational manual carefully before using the machine.
2.         The centrifuge’s working environment: temperature: 5-40℃, relative humidity ≤80%. No electric dust, explosive gas, corrosive gases in the air.
3.         Put the machine on a level and strong table and keep the working room clean, dry and ventilated.
4.         Clean the rotor and container before operating.
5.         Place empty centrifugal tube evenly before working, after working 1-2 minutes at the maximum speed, then begin to work if there’s no problem.
 
Using Description
1.       Prepare according to before using should know.
2.       Put the test tube (with same weight) into the centrifugal hole. Close the lid.
3.       Power on, and press switch button, turn on the machine.
4.       Press the “switch” knob, choose the timer, by press “+” “-” to setup the time you need.
5.       Press the “switch” knob, choose the speed, by press “+” “-” to setup the speed you need. You can choose 1000r/min, 1500r/min, 2000r/min, 2500r/min, 3000r/min, 3500r/min, and 4000r/min.
6.       After using, turn off the power, clean the machine.