Usage
This machine is used to photograph patient’ s esophagus, chest, gastrointestinal, abdomen, limbs etc for common radiography, Beam Limiting Device radiograph and high KV radiography.
Features
1. The console applies LCD display photograph term, the device have the mA, kV, s of photograph table projection chosen are controlled by procedure interlock, be separate , not disturb.
2. The console control the expose time by mcu internal clock, make the expose time accurate.
3. This machine adopts mA, KV, s three parameters free matching work method. According to x-ray capacity constrains, by software for producing interlock protection conditions.
4. The low-voltage power supply was isolated in this machine, in particular, MCU power supply; fully protect the stability of the work of the MCU.
5. Expose mode of this machine use two-step hand-swicth control mod, hand swicth control circuit adopt a safe voltage control.
6. The conversion of diagnosis table and photography table is finished by slotter key which is on the pane.