 [Main technical parameters]
Measuring range
CO: 0~10.0 10-2(%)    CO2: 0~20.0 10-2(%)
HC: 0~10000 10-6(ppm)    O2: 0~25.0 10-2(%)
Oil temperature: 18℃~150℃ 
Speed: 100~20000rpm
Repeatability error
≤±2%
Zero drift
CO: ±0.06 10-2(%)/4h   CO2: ±0.5 10-2(%)/4h
HC: ±30 10-6(ppm)/4h
O2: ±0.5 10-2(%)/4h      NO: ±30 10-6(ppm)/4h
Span drift
No more than indication error allowable value (4 hours)
Indication error
CO: ±5%(Relative error, 0~10%)
CO2: ±5%(Relative error, 0~20%)
HC: ±5%(Relative error, 0~4000ppm)±10%(Relative error, 4001~10000ppm)
O2: ±5%(Relative error, 0~25%)
NO: ±5%(Relative error, 0~5000ppm)
Environmental conditions
oil temperature: ± 2 ℃ (relative error)
Speed: ± 1.0% (relative error)
Temperature: 0 ℃ ~ 40 ℃
 Humidity: ≤ 80%
Atmospheric pressure: 60 ~ 105KPa
 Power: AC 220V ± 10%; 50Hz ± 1Hz
Preheating time
10min
Weight
15KG
Volume
610 * 500 * 330mm
Simpling  Exhaust
direct sampling, the length of the sampling tube is 5m, the length of the sampling probe is 900mm
Oil temperature: temperature measurement inserts into the oil gauge hole of the engine, insert length is same with the oil gauge, blocked with rubber plug to prevent oil smoke
Wire length: 5m;
Speed: fix the folder of tachometer on the high-voltage soft-line the of gasoline engine to measure the speed.
Display Mode 
High-brightness LED display    Large LCD display
Optional configuration
 temperature, speed configuration
Main functions
 Printer built in(temperature, speed configuration) , Communication with the computer network, 30 minutes automatically prompte to zero. Idling and dual idling two measurement methods, built-in clock module to display time, number plates can be stored 500 sets of measurements data, natural gas, liquefied petroleum gas and gasoline these three fuels to choose.