1. Features
◆ 3 channel ECG machine with interpretation
◆ Manual/Auto/Arrhythmia Analysis Modes
◆ 80mm 3-ch recording with rhythm lead
◆ Built in rechargeable Li-ion battery
◆ Adapter 110-230V, 50/60Hz AC power supply
◆ Built in USB/RS232 interface(ECGNET software optional)
◆ Manual/Auto/Arrhythmia Analysis Modes
◆ 80mm 3-ch recording with rhythm lead
◆ Built in rechargeable Li-ion battery
◆ Adapter 110-230V, 50/60Hz AC power supply
◆ Built in USB/RS232 interface(ECGNET software optional)
2. Technical Specification
◆ Lead: standard 12 leads
◆ Lead acquisition: 12bit/1000Hz(synchronously 12 leads)
◆ Work modes: MAN, AUTO, ANALYSIS modes
◆ Filter:
AC Flter : 50Hz / 60Hz
EMG Filter : 25Hz / 45Hz
Anti-Drift Filter : 0.15Hz (Adaptive)
◆ CMMR: ≥120dB (with AC filter)
◆ Input circuit: floating; protection circuit against defibrillator effect
◆ Input Impedance: ≥50MΩ
◆ Input circuit current: ≤0.05μA
◆ Input dynamic range: ±10Vpp
◆ Patient current leakage: <10μA
◆ Calibrating voltage: 1mV±2%
◆ Voltage tolerance: ≥±500mV
◆ Time constant: ≥3.2s
◆ Frequency response: 0.05~160Hz(-3db)
◆ Noise level: <15μVp-p
◆ Sensitivity valve: ≤20μV
◆ Sensitivity: 5, 10, 20mm/mV
◆ Inter-channel Interference: ≤0.5mm
◆ Paper speed: 25, 50mm/s(±3%)
◆ Display:
ECG-9031: 320x240 graphic 3.8 inch LCD
◆ Dimension and weight: 285x265x72(mm), about 2.5Kg
3. Standard Configuration
◆ Patient cable 1 set
◆ Limb electrode 4 sets
◆ Chest elelctrodes 6 sets
◆ Grounding cable 1 set
◆ Paper axis 1 set
◆ Thermal paper 1 set
◆ Fuse 2 sets
◆ 3-cord power cord 1 set
◆ AC Adaptor 1 set