RT3303B Portable Three Phase Energy Meter Test Instrument
Overview:
The RT3303 Series Portable Three Phase Energy Meter Test Instrument comprises of high-stability three phase program-controlled power source and three phase standard watt-hour meter.
It applies to the calibration of electronic and electro-mechanical single phase and three phase active and reactive and reactive watt-hour meters as well as multi-function watt-hour meters.
It can provide us with rich electric quantity information such as AC voltage, AC current, phase difference between voltage and current, power factor, active power, reactive power, apparent power, frequency, and so on.
It can be hung with 1 to 3 meter stations. It is embedded with error processor or can be inserted with error processor.
It can be used in such area as product research and development, product maintenance, and product test at production lines. By using the underlying dynamic link library developed by Rowwin Test, users may develop their own automatic meter calibration software.
The RT3303 Series Portable Three Phase Energy Meter Test Instrument comprises of high-stability three phase program-controlled power source and three phase standard watt-hour meter.
It applies to the calibration of electronic and electro-mechanical single phase and three phase active and reactive and reactive watt-hour meters as well as multi-function watt-hour meters.
It can provide us with rich electric quantity information such as AC voltage, AC current, phase difference between voltage and current, power factor, active power, reactive power, apparent power, frequency, and so on.
It can be hung with 1 to 3 meter stations. It is embedded with error processor or can be inserted with error processor.
It can be used in such area as product research and development, product maintenance, and product test at production lines. By using the underlying dynamic link library developed by Rowwin Test, users may develop their own automatic meter calibration software.
Technical Parameters:
Voltage and Current Measurement Ranges Output Capability
Output Voltages 57.7V/63.5V,220V/240V
(480V also available) Voltage Loop Max.50VA
Output Currents 0.1A,1A,5A,20A,100A Current
Loop Max.100VA
(For measurement range of 100A)
Voltage and Current Measurement Ranges Output Capability
Output Voltages 57.7V/63.5V,220V/240V
(480V also available) Voltage Loop Max.50VA
Output Currents 0.1A,1A,5A,20A,100A Current
Loop Max.100VA
(For measurement range of 100A)
Output Adjustment Adjustment Range Adjustment Fineness Max. Output
Voltage/Current
Voltage 0% to 125%
Each shift value 10%,1%,0.1% and 0.01% 300V/600V
Current 0% to 120%
Each shift value 10%,1%,0.1% and 0.01% 120A
Phase 0 to 359.99 10, 1, 0.1 and 0.01
Voltage/Current
Voltage 0% to 125%
Each shift value 10%,1%,0.1% and 0.01% 300V/600V
Current 0% to 120%
Each shift value 10%,1%,0.1% and 0.01% 120A
Phase 0 to 359.99 10, 1, 0.1 and 0.01
1. Accuracy
Class 0.1 (independent standard reference energy meter RT5303)
2. Output voltage and current distortion
Output voltage and current distortion not larger than 0.3% (if it is a linear load)
3. Output power stability
The variation of pre-set output power within 180S will not exceed 0.02% compared with the measurement range
4. Electric quantity symmetry among the three phases
The difference between any line voltage or phase voltage and its mean value will not be larger than 0.2%; the difference between any phase current and its mean value will not be larger than 0.2%; the phase difference between any phase current and its corresponding phase voltage will not be larger than 0.2 degree
5. Harmonic wave setting
a. Number of harmonic wave: 2 to 31
b. Content of harmonic wave: less than 40% (compared with base wave)
c. Phase of harmonic wave: 0 to 359.99
6. Monitoring instrument
a. (Embedded standard reference energy meter) Monitoring instrument’s accuracy class and resolution:
Current and Voltage: 0.1%;
Phase: 0.1 degree;
Frequency: 0.01 Hz
b. (Insert-type standard reference energy meter) Reading will be made directly from the standard reference energy meter
7. Power supply
AC single-phase 220V/240V (100V/110V also available), 50/60Hz, maximum input power consumption: 500W
Class 0.1 (independent standard reference energy meter RT5303)
2. Output voltage and current distortion
Output voltage and current distortion not larger than 0.3% (if it is a linear load)
3. Output power stability
The variation of pre-set output power within 180S will not exceed 0.02% compared with the measurement range
4. Electric quantity symmetry among the three phases
The difference between any line voltage or phase voltage and its mean value will not be larger than 0.2%; the difference between any phase current and its mean value will not be larger than 0.2%; the phase difference between any phase current and its corresponding phase voltage will not be larger than 0.2 degree
5. Harmonic wave setting
a. Number of harmonic wave: 2 to 31
b. Content of harmonic wave: less than 40% (compared with base wave)
c. Phase of harmonic wave: 0 to 359.99
6. Monitoring instrument
a. (Embedded standard reference energy meter) Monitoring instrument’s accuracy class and resolution:
Current and Voltage: 0.1%;
Phase: 0.1 degree;
Frequency: 0.01 Hz
b. (Insert-type standard reference energy meter) Reading will be made directly from the standard reference energy meter
7. Power supply
AC single-phase 220V/240V (100V/110V also available), 50/60Hz, maximum input power consumption: 500W