Introduction of product:
DK-34B1 Multi-function AC sampling and Transmitter Calibrating Instrument adopts advanced technology of large-scale import integrated circuit, high-speed digital signal processor (DSP), DDS waveform synthesis, complex programmable logic device, large-scale integrated function, original operating system, etc. to design. It is true of testing and calibrating AC sample, Transmitter and indicating instrument, the ideal equipment for industrial and mining enterprises and power system that be used to test and calibrate products.
Main Features:
●    Adopt the design of broadband standard source, large power output, high stability, fast response speed.
●    The first in China to integrate system, test and signal generate into a module, high integration, low fault rate, small volume, light weight, high stability.
●    Adopt original operating system, starting up show the test image immediately, output of the standard source, no need bootstrap, fast response speed, high work efficiency.
●    Voltage, current, power, phrase, frequency are all adopt the output of excellent closed-loop system, software adjustment, one-time set point, easy to use.
●    Voltage, current, phrase set abundant common experiment point, easy to operate, one-click to complete, convenient and efficient to use.
●    Equipped with digital rotary encoder adjustment, convenient and easy to use.
●    Application of electric power communication protocol and computer, verification device can be conveniently carry on fully-automatic testing and managing to AC sampling devices and transmitters.
●    Powerful and flexible software function, the first in China to equip with template protocol and report, allow customers to define new RTU communication protocol and new report format by template, software provide factories with nearly 100 kinds of communication protocol template of common RTU model, and equip with powerful online help and voice prompt device.
●    Clamp meter and standard source integrative structure, the front panel is standard source output, the back panel is attached standard input, verify of protocol and on-spot (equip with the clamp meter) that function two for one.
●    Output of standard harmonic 2~31 times, processing the output of single and random superimposed multiple harmonic.
●    It can randomly shift phrase between three-phase voltage, three-phase current and per phase voltage and current, thus can simulate the output of various power failure also.
●    It has the function of multiple alarming and protecting, fault detection automatically, and showing the fault type and position, safe and reliable to use.
●    Equipped with interface and software, the open interface protocol, customers can individually program and control apparatus.
 
Main specification:
 
Main specification
Output AC Voltage
57.5V, 100V, 220V, 380V, 0~120% adjustable, finess:0.002%  
Accuracy:0.005%
Output AC
1A, 2A, 5A, 20A, 0~120% adjustable, finess:0.002%
Accuracy:0.005%
Output DV
75mV, 10V, 100V, 300V, 600V, 0~120% adjustable, finess:0.002%
Accuracy:0.005%
Output DC
20mA, 10mA, 1mA
Accuracy:0.005%
DC measurement
10V, 5V, 2.5V, 20mA, 10mA, 5mA, 1mA
range: 0~120% adjustable
Stability:0.01%/2min
Output AC Voltage,
AC current
Distortion≤0.2%
Accuracy:0.005%
Output AC Voltage,
current load
per phase voltage:35VA, per phase current: 25VA
Load regulation
<0.02%
Output active/
reactive power
The accuracy of active power:0.05%, the accuracy of reactive power: 0.1%
 
Output frequency
45Hz~65Hz, finess:0.001Hz
Accuracy:0.002Hz
Output phrase
0~359.99°, finess:0.01°
Accuracy:0.05°
Output power factor
-1~0~+1, finess:0.0001
Accuracy:0.05%
Output harmonic
2~20times, can multiple superimposed,
0~40% adjustable
Accuracy:2%
Output harmonic
21~31times, can multiple superimposed,
0~40% adjustable
Accuracy:5%
Input voltage
measurement
57.5V, 100V, 220V, 380V, range switch automatically
Accuracy:0.05%
Input current
measurement
5A, range(0~120% adjustable)
Accuracy:0.05%
Power measurement
The accuracy of active power:0.05%, the accuracy of reactive power:0.1%
 
Clamp meter
measurement
Measure range:5A
Accuracy:0.2%
With RS232 interface protocol
Can complete RS232 interface protocol
 
Volume
450×440×132(mm)
 
Power supply
Single-phase 220V±10%
 
Weight
About 15kg