YC1891D single phase electric energy meter test system can measure single phase mechanical meter, electric mechanical meter and electronic meter by automatic, semi-automatic or Manual. The material is aluminum alloy, light and strong, prevent corrosion.
Function:
- Starting test and creeping test
- Measuring the different constant meters synchronously.
- With the protect function of overload, short voltage circuit and open current circuit.
- The testing software with friendly Windows 2000 and XP interface, include functional module of all kinds of modern energy meter needed, to reality error test, harmonic test, demand test, influence test.
- The stored testing data can be transferred to multi-format for other program using
Character:
- Meter under testing: 1P2W
- Testing mode: active power, reactive power
- With MSVT, can test single-phase meters with closed links between the current and voltage measuring circuits (I-P links)
- With current output switch, can calibrate 2 element meter (optional)
- Installing RS232, Ethernet communication port in each meter position, to reality the automatic calibrate of electronic meter (optional)
Technical data:
|
|
YC1891D-6 |
YC1891D-12 |
YC1891D-24 |
YC1891D-48 |
|
Number of meter positions |
6 |
12 |
24 |
48 |
|
Accuracy |
0.05 |
|||
|
Voltage |
||||
|
Test voltage output |
24V – 288V or customized |
|||
|
Power of the voltage output |
100VA |
250VA |
500VA |
1000VA |
|
Resolution |
Better than 0.01% of full scale value of range |
|||
|
Setting accuracy |
Better than 0.05% at the final range value |
|||
|
Stability |
Better than 0.005%/h (integration time 150 s) |
|||
|
Load regulation from 0-max. load |
Better than 0.01% |
|||
|
Distortion factor |
Better than 0.3% for linear resistance load |
|||
|
Harmonics |
2nd – 21st free programmable |
|||
|
Current |
||||
|
Test current output |
1mA – 100A or customized, Max.200A |
1mA – 100A |
||
|
Power of the current output |
150VA |
350VA |
750VA |
1500VA |
|
Resolution |
Better than 0.01% of full scale value of range |
|||
|
Setting accuracy |
Better than 0.05% at the final range value |
|||
|
Stability |
Better than 0.005%/h (integration time 150 s) |
|||
|
Load regulation from 0-max. load |
Better than 0.01% |
|||
|
Distortion factor |
Better than 0.3% for linear resistance load |
|||
|
Harmonics |
2nd – 21st free programmable |
|||
|
Phase angle |
||||
|
Range |
0 -- 360° |
|||
|
Resolution |
0.01° |
|||
|
Setting accuracy |
0.1° |
|||
|
Frequency |
||||
|
Range |
45Hz – 65Hz |
|||
|
Resolution |
0.01Hz |
|||
|
Error display |
||||
|
Type of error display |
Red LED |
|||
|
Resolution of the error displays |
8 digits |
|||
|
Others |
||||
|
Voltage Supply |
220V±10% or customized 50/60Hz±2Hz |
3x220/380V±10% or customized 50/60Hz±2Hz |
||
|
Ambient temperature |
-10°C -- +40°C |
|||
|
Relative humidity |
35% -- 85% |
|||

