l Using mixed gas containing 5% Hydrogen and 95% Nitrogen as a tracer gas
l Non flammable safe gas approved by ISO10156 : 1990
l Minimum detectable leak rate of 5×10-8Pa.m3/s
l Accurate positioning with high sensitivity
l Auto conversion among ppm, g/y, ml/min and Pa.m3/s
Testing Method
|
Ciffusion Type
|
Mounting Type
|
Portable, or panel mounting
|
Sensitivity
|
Leak Rate: 5×10-8 Pa.m3/s(5%H2),Concentration: 0.5 ppm
|
Leakage Range
|
1×10-4 ~5×10-8 Pa.m3/s(5%H2)
|
Concentration Detection Range
|
0.5 ~ 1000 ppm
|
Concentration Calibration Accuracy
|
20% F.S.
|
|
|
Display
|
LCD Display (Blue LED Backlight)
Dot Matrix: 320×240
W124mm×D94mm .
Effective Display-Size : W124mm×D94mm .
|
Group Setting
|
0~15ch(16ch Total)
|
External Terminals
|
Input Signals: 10
Output Signals: 8
Valve Control Signals:8
|
Power Source
|
AC 100~240V, 50/60Hz
|
Power Consumption
|
< 50W
|
Operation
Temperature/Humidity Range
|
5~40 ℃, 45~85%RH, With no Precipitation
|
Weight
|
Main Body: 12KG
|