Product Main

Specifications

Standard models with full encodor switch
TOS-2020B / 2040B / 2050B

TOS-2020BF / 2040BF / 2050BF: with built-in 6 digits frequency counter

 
Features

Dual channel 20MHz/40MHz/50MHz

High sensitivity 1mV/DIV
Z-axis modulation input, CH1 signal output

10 times sweep magnification

TV synchronization, X-Y mode

High illumination internal graticule CRT

Encoder for sweep and vertical mode switch

Fully sealed durable attenuation switch

ALT triggering function, enabling simultaneous observation of two independent signals

Trigger level lock function, Auto synchronize function

Built-in 1000:1 attenuator, expanded vertical sensitivity to 10V~20V/DIV

 
 
Specifications
MODEL
TOS-2020B
TOS-2020BF
TOS-2040B
TOS-2040BF
TOS-2050B
TOS-2050BF
Vertical system
Sensitivity

5mV~20V/DIV, 12 steps in 1-2-5 sequence

Sensitivity accuracy

≤3%; X5MAG: ≤5%

Vernier sensitivity
1/2.5 or less than panel indicated value
Bandwidth
DC~20MHz

(X5MAG: DC~7MHz)

DC~40MHz

(X5MAG: DC~15MHz)

DC~50MHz

(X5MAG: DC~15MHz)

AC coupling

<10Hz (at 100kHz, 8DIV, frequency response -3dB)

Rise time
Approx.17.5ns
(X5MAG: Approx.50ns)
Approx.8.75ns
(X5MAG: Approx.25ns)
Approx.7ns
(X5MAG: Approx.23.3ns)
Input impedance
Approx.1MΩ//Approx.25pF

DC balance shift

5mV~5V/DIV: ±0.5DIV

Vertical mode

CH1, CH2, DUAL (ALT/CHOP), ADD, CH2 INV

Chopping repetition frequency

Approx. 250kHz

Input coupling

AC, GND, DC

Max. Input voltage

300V peak-peak (AC: frequency≤1kHz)

Common mode rejection ratio

>50:1 at 50kHz sine wave (set CH1 and CH2 at same sensitivity)

CH2 INV BAL

Balanced point variation≤1DIV (referred to graticule center)

CH1 signal output

Min.20mV/DIV (50Ω output bandwidth 50Hz~5MHz)

 
 
MODEL
TOS-2020B
TOS-2020BF
TOS-2040B
TOS-2040BF
TOS-2050B
TOS-2050BF
Horizontal system
Sweep time base
0.2us~0.5s/DIV, 20 steps in 1-2-5 sequence

Sweep time base accuracy

±3%, X10MAG: ±5% (20ns~50ns/DIV uncalibrated)

Vernier sweep time base

≤1/2.5 of panel indicated value

Sweep magnification

X10 (Max. sweep time base 20ns/DIV)

Position shift caused by X10MAG

Within 2DIV at CRT screen center

Linearity

±5%, X10MAG: ±10% (0.2s~1us)

Trigger system
Trigger mode
AUTO, NORM, TV-V, TV-H
Trigger level lock
Provided

Trigger source

CH1, CH2, LINE, EXT

Trigger coupling

AC: 20Hz to full bandwidth

Polarity

“+” or “-” polarity

Sensitivity
 
20Hz~2MHz
2MHz~20MHz
20MHz~40MHz
40MHz~50MHz
CH1, CH2
1DIV
1.5DIV
2.5DIV
3DIV
ALT
2DIV
3DIV
 
 
EXT
200mV
800mV
 
 
TV: Sync pulse>1DIV (EXT: 1V)

External trigger output signal

Input impedance: Approx.1MΩ//Approx.25pF
Max. input voltage: 300V (DC+AC peak), AC≤1kHz
Z-axis
Sensitivity
5Vp-p
Bandwidth
DC~2MHz
Input impedance
Approx.47kΩ
Max. input voltage

30V (DC+AC peak, AC frequency≤1kHz)

Calibration signal
Waveform

Positive-going square wave

Frequency
Approx.1kHz
Duty ratio
<48:52

Output voltage

2Vp-p ±2%

Output impedance

Approx.1kΩ
X-Y mode
Sensitivity
5mV~5V/DIV, ±3%
Bandwidth

DC~500kHz (-3dB)

X-Y phase error

≤3°at DC~50kHz

Frequency counter (only for TOS-2020BF, TOS-2040BF, TOS-2050BF)
Display
6 digits LED
Display accuracy
0.01%
CRT
Mode

6" rectangle, internal graticule

Display area
8×10DIV (1DIV=1cm)
Phosphor
P31

Accelerating voltage

Approx.2kV (20MHz); Approx.12kV (40MHz)

Trace rotation

Adjustable at front panel
General
Power source

AC110V/220V±10%, 50/60Hz, Max.35VA

Dimension
310×150×455mm
Accessories

Power cord ×1, Operation manual ×1, Probe ×2

Weight
Approx.8kg