Main Features:
4.3inch 480x272 color graphic LCD display
Three channel ECG machine with interpretation
12-lead ECG simultaneous acquisition
Multiple operation modes: Manual/Auto/Arrhythmia Analysis Modes
80mm 3-channel recording with a rhythm lead
Digital filters eliminate EMG, AC, Baseline Drift interference
Built in rechargeable Li-Ion battery, 3 hours continuous working
Adapt 110-240V, 50/60Hz AC power supply
250 patients save/replay
Built in USB/RS232 interface
4.3inch 480x272 color graphic LCD display
Three channel ECG machine with interpretation
12-lead ECG simultaneous acquisition
Multiple operation modes: Manual/Auto/Arrhythmia Analysis Modes
80mm 3-channel recording with a rhythm lead
Digital filters eliminate EMG, AC, Baseline Drift interference
Built in rechargeable Li-Ion battery, 3 hours continuous working
Adapt 110-240V, 50/60Hz AC power supply
250 patients save/replay
Built in USB/RS232 interface
Lead:
|
Standard or Cabrera 12 leads,
Simultaneously 12 leads acquisition
|
Input Circuit:
|
Isolated as well as pace make and defibrillation protected
|
Patient Current Leakage:
|
<10mA
|
Input Impedance:
|
>50MW
|
Frequency Response:
|
0.05 ~ 150Hz
|
Time Constant:
|
³ 3.2s
|
CMRR:
|
> 80dB; > 100dB (filter in use)
|
EMG Filter:
|
25Hz (-3dB) or 35Hz (-3dB)
|
Sensitivity:
|
5, 10, 20mm/mV±5%
|
Recording Mode:
|
Thermal array printer
8 dot/mm(Vertical) 40 dot/mm(Horizontal, 25mm/s)
|
Paper Speed:
|
5, 10, 25, 50mm/s ± 5%
|
Paper Size:
|
80mm´15m, roll type
|
Safety Level:
|
Class I, type CF
|
Power Requirements:
|
AC: 220V±10%, 50Hz±5%, 50VA
Built-in Li battery, 12V
|
Dimension:
|
285mm×200mm×50mm
|
Weight:
|
2.5Kg
|