Product Main

Specifications

Features

♦ High-resolution thermal printing array system

♦ Frequency response is as high as 150Hz

♦ Capable of printing continuously on channel trace and annotations including lead mark and parameter such 
   as sensitivity, paper speed and filter operation status

♦ High resolution digital filter inhibits baseline drift without affecting ECG waveform

♦ One-touch operation

♦ Concept of floating input circuit meets IEC safety standard

♦ Rechargeable battery supporting more than 50 patient examinations 

♦ Lead Standard: 12 leads


Specification:

♦ Input: floating, protection against defibrillator effect

♦ Lead: standard 12 leads, lead change automatically

♦ Patient current leakage: <10μA

♦ Input impedance: ≥10MΩ

♦ Calibrating voltage: 1mV±3%

♦ A/D conversion: 12 bit

♦ Frequency response: 0.05Hz~150Hz

♦ Time constant: >3.2 seconds

♦ CMRR: >80dB, >100dB (with filter)

♦ EMG filter: 35Hz (-3dB) / 25Hz (-3dB)

♦ Sensitivity: 1/2, 1, 2 (cm/mV), conversion deviation≤5%

♦ Recording system: thermal-array, 8 dots / mm (vertical), 16 dots / mm (horizontal, 25mm / sec)

♦ Paper speed: 25, 50 mm / s ± 3%

♦ Recording paper: 50mm, 20m, high-speed roll

♦ Input circuit current: ≤50nA

♦ Safety standard: IEC classΙ, type CF

♦ Power supply: AC: 20V/110V, 50/60 Hz;
                             DC: 12V rechargeable battery

♦ Skin voltage tolerance: ±300mV

♦ Noise level: <15μVp-p

♦ Dimensions: 310×230×70mm

♦ Dimension of packing: 400×280×160mm

♦ Weight: 2.75KG


Standard configuration:

• 1×ECG machine
• 1×patient cable
• 4× limb electrodes
• 6×chest electrodes
• 1×power cable
• 1×grounding cable
• 1×printing paper
• 1×operation manual
• 1×packing list