8.4 inch SPO2/ECG/NIBP Patient Monitor
Features:
Features:
1. 8.4 inch color TFT display
2. Unique digital SPO technology with excellent measurement and interference-free
3. Audible and visual alarms with adjustable alarm range
5. Anti electrosurgical interference and defibrillation
6. Can be connected with LAN to work with central monitoring system
7. Powerful data management and storage capacity
8. Suitable for adult, pediatric and neonatal patient
Standard configuration:
ECG, NIBP, SPO2, PR, Built-in rechargeable battery
Optional:
2. Unique digital SPO technology with excellent measurement and interference-free
3. Audible and visual alarms with adjustable alarm range
5. Anti electrosurgical interference and defibrillation
6. Can be connected with LAN to work with central monitoring system
7. Powerful data management and storage capacity
8. Suitable for adult, pediatric and neonatal patient
Standard configuration:
ECG, NIBP, SPO2, PR, Built-in rechargeable battery
Optional:
RESP, TEMP, 2-TEMP, 2-IBP, ETCO2, Thermal printer
Technical Specifications:
Technical Specifications:
ECG:
1. Lead Mode: 5 Leads (R, L, F, N, C or RA, LA, LL, RL, V)
2. Lead selection: I, II, III, avR, avL, avF, V
3. Waveform: 2 ch
4. Gain: 2.5mm/mV, 5.0mm/mV, 10mm/mV, 20mm/mV
5. HR and Alarm Range:
1) Adult: 15 ~ 300 bpm
2) Neo/Ped: 15 ~ 350 bpm
3) Accuracy: ±1% or ±1bpm, which great
4) Resolution: 1 bpm
5) Sensitivity: >200uV P-P
6. Differential Input Impedance: >5MΩ
7. CMRR:
1) Monitor: >105dB
2) Operation: >105dB
3) Diagnosis: >85dB
8. Electrode offset potential: 300mV
9. Leakage Current: <10uA
10. Baseline Recovery: <3s after Defi.
11. ECG Signal Range: 8mV(Vp-p)
12. Calibration Signal: 1mV (Vp-p), 5% Accuracy
13. ST Segment measure and alarm: -2.0 ~ +2.0 mV
NIBP:
2. Lead selection: I, II, III, avR, avL, avF, V
3. Waveform: 2 ch
4. Gain: 2.5mm/mV, 5.0mm/mV, 10mm/mV, 20mm/mV
5. HR and Alarm Range:
1) Adult: 15 ~ 300 bpm
2) Neo/Ped: 15 ~ 350 bpm
3) Accuracy: ±1% or ±1bpm, which great
4) Resolution: 1 bpm
5) Sensitivity: >200uV P-P
6. Differential Input Impedance: >5MΩ
7. CMRR:
1) Monitor: >105dB
2) Operation: >105dB
3) Diagnosis: >85dB
8. Electrode offset potential: 300mV
9. Leakage Current: <10uA
10. Baseline Recovery: <3s after Defi.
11. ECG Signal Range: 8mV(Vp-p)
12. Calibration Signal: 1mV (Vp-p), 5% Accuracy
13. ST Segment measure and alarm: -2.0 ~ +2.0 mV
NIBP:
1. Method: Oscillometry
2. Mode: Manual, Auto, STAT
3. Measuring Interval in AUTO Mode: 1, 2, 3, 4, 5, 10, 15, 30, 60, 90Min
4. Measuring Period in STAT Mode: 5 Min
5. Pulse Rate Range: 40 ~ 240 bpm
6. Alarm Type: SYS, DIA, MEAN
7. Measuring and alarm Range
1) Adult Mode:
SYS: 40 ~ 270 mmHg
DIA: 10 ~ 215 mmHg
MEAN: 20 ~ 235 mmHg
2) Pediatric Mode:
SYS: 40 ~ 200 mmHg
DIA: 10 ~ 150 mmHg
MEAN: 20 ~ 165 mmHg
3) Neonatal Mode:
SYS: 40 ~ 135 mmHg
DIA: 10 ~ 100 mmHg
MEAN: 20 ~ 110 mmHg
Resolution: 1mmHg
Over-pressure protection: YES
SPO2:
2. Mode: Manual, Auto, STAT
3. Measuring Interval in AUTO Mode: 1, 2, 3, 4, 5, 10, 15, 30, 60, 90Min
4. Measuring Period in STAT Mode: 5 Min
5. Pulse Rate Range: 40 ~ 240 bpm
6. Alarm Type: SYS, DIA, MEAN
7. Measuring and alarm Range
1) Adult Mode:
SYS: 40 ~ 270 mmHg
DIA: 10 ~ 215 mmHg
MEAN: 20 ~ 235 mmHg
2) Pediatric Mode:
SYS: 40 ~ 200 mmHg
DIA: 10 ~ 150 mmHg
MEAN: 20 ~ 165 mmHg
3) Neonatal Mode:
SYS: 40 ~ 135 mmHg
DIA: 10 ~ 100 mmHg
MEAN: 20 ~ 110 mmHg
Resolution: 1mmHg
Over-pressure protection: YES
SPO2:
1. Measuring Range: 0 ~ 100%
2. Alarm Range: 0 ~ 100%
3. Resolution: 1%
4. Accuracy:
1) 70% ~ 100%: 2 %
2) 0% ~ 69%: unspecified
5. Actualization interval: About 1Sec.
6. Alarm Delay: 10 Sec.
7. Pulse Rate:
1) Measuring and Alarm Range: 0~254bpm
2) Resolution: 1bpm
3) Accuracy: 2bpm
4) Alarm Delay: 10 Sec.
2. Alarm Range: 0 ~ 100%
3. Resolution: 1%
4. Accuracy:
1) 70% ~ 100%: 2 %
2) 0% ~ 69%: unspecified
5. Actualization interval: About 1Sec.
6. Alarm Delay: 10 Sec.
7. Pulse Rate:
1) Measuring and Alarm Range: 0~254bpm
2) Resolution: 1bpm
3) Accuracy: 2bpm
4) Alarm Delay: 10 Sec.