Product Main

Specifications

Summary:
Used for testing the content of the released halogen acid gas, pH value (hydrogen ion content), conductivity etc. when the cable is burning.
Standard: GB/T 17650 IEC 60754- 1 and IEC 60754-2.

Technical parameters:
1. Activated carbon (air filter) : filter air source through the filter
2. Silica gel (air dry) : a device for drying air source
3. Flowmeter : 1L/min, a device for controlling the flow of the air source
4. Thermocouple : K type of high temperature resistant, thermocouple stainless, 0 ~ 1300℃
5. Conductivity: ¢40*800
6. Burning ship mobile system :a device for moving the burning ship automatically.
7. Drawer : a box used for storing fittings (such as conductivity meter, pH gauge etc)
8.Magnetic stirrers : an stirrer used for measuring the burning gas of electrical conductivity of the (ray magnetic)
9. PH meter : a device used for measuring the pH acidity (ray magnetic)
10. Conductometer : an instrument used for measuring the conductivity of the burning gas (ray magnetic)
11. 7 "color touch screen and PLC and PID control heating furnace temperature automatic calculation module
12. Heating furnace : ¢220*600, Effective heating space: ¢43*550, power: 4KW
13. Washing bottle : 500ml, 2 Straight-pipe washing bottles, 1 measuring pot of 2L
14. Installation size : 1600mm(W)*600(D)
15. Power source : voltage: 220V,  current:21A