Specifications

Technical Specification of Anesthesia Machine         

 

Physical & Environmental Specification         

  • Dimensions(W×H×D): 890 ×1335×620mm         
  • Anesthesia Machine Weight: 110kg         
  • Storage temperature: -20~+55℃         
  • Working temperature: +10~+40℃         
  • Storage humidity: ≤93%         

Electronical Specification         

  • Mains: 110-240V AC, 50-60Hz         
  • Power consumption: ≤50W         

Pneumatics Specification         

  • Gas supply: O2,N2O: 0.28-0.6Mpa         
  • Flowmeters: O2: 0-1L/min; 1-10L/min; N2O: 0-1L/min; 1-10L/min         
  • Gas system: O2 deficiency alarm; Hypoxic guard system        ; O2 flush: 35-75L/min         
  • Working mode:         Closed, Semi-closed, Semi-open         
  • Driven mode: Pneumatically driven and electronically controlled         
  • Operating mode: Man. / Vent.         
  • Safety valve: ≤8kpa         

Ventilator Specification         

  • Patient type: Adult,pediatric         
  • Ventilation modes: IPPV, Manual         
  • Setting: Touch key         
  • Tidal volume: 0-1500ml         
  • Ventilation frequency: 6-60bpm         
  • I:E ratio: 2:1-1:4         
  • Ppeep:         0-20 cmH2O         
  • High pressure: 21-60cmH2O         
  • Low pressure: 0-20cmH2O         
  • Monitoring: Display: High quality LED        
          Pressure: Paw        
          Volumes: Vt        

          Breath rate         

  • Alarm:         Pressure: Paw high/low limit        

    Low O2 supply pressure         

    Power failure         

    Alarm silence: ≤120s         

Vaporizer                 

  • Support 1 vaporizer (Selectatec compatible )         
  • Agent type: Halothane, Enflurane, Isoflurane , Sevoflurane available.