Technical Specification of Anesthesia Machine
Physical & Environmental Specification
- Dimensions(W×H×D): 890 ×1335×620mm
- Anesthesia Machine Weight: 110kg
- Storage temperature: -20~+55℃
- Working temperature: +10~+40℃
- Storage humidity: ≤93%
Electronical Specification
- Mains: 110-240V AC, 50-60Hz
- Power consumption: ≤50W
Pneumatics Specification
- Gas supply: O2,N2O: 0.28-0.6Mpa
- Flowmeters: O2: 0-1L/min; 1-10L/min; N2O: 0-1L/min; 1-10L/min
- Gas system: O2 deficiency alarm; Hypoxic guard system ; O2 flush: 35-75L/min
- Working mode: Closed, Semi-closed, Semi-open
- Driven mode: Pneumatically driven and electronically controlled
- Operating mode: Man. / Vent.
- Safety valve: ≤8kpa
Ventilator Specification
- Patient type: Adult,pediatric
- Ventilation modes: IPPV, Manual
- Setting: Touch key
- Tidal volume: 0-1500ml
- Ventilation frequency: 6-60bpm
- I:E ratio: 2:1-1:4
- Ppeep: 0-20 cmH2O
- High pressure: 21-60cmH2O
- Low pressure: 0-20cmH2O
- Monitoring: Display: High quality LED
Pressure: Paw
Volumes: Vt
Breath rate
- Alarm: Pressure: Paw high/low limit
Low O2 supply pressure
Power failure
Alarm silence: ≤120s
Vaporizer
- Support 1 vaporizer (Selectatec compatible )
- Agent type: Halothane, Enflurane, Isoflurane , Sevoflurane available.