Anaesthesia Machine Technical Specification
Physical & Environmental Specification
- Dimensions(W×H×D): 930×1354×620mm
- Anaesthesia Machine Weight: 120kg
- Storage temperature: -10~+55℃
- Working temperature: +10~+40℃
- Storage humidity: ≤93%
Electronical Specification
- Main power: 110-240V AC, 50-60Hz
- Battery(option): 24V DC, minimum 90 min
- Power consumption: ≤70W
Pneumatics Specification
- O2: 0-1L/min; 1-10L/min
- N2O: 0-1L/min; 1-10L/min
- Gas system: O2 deficiency alarm
- Hypoxic guard system
- O2 flush: 35-75L/min
- Working mode: Closed, Semi-closed, Semi-open
- Driven mode: Pneumatically driven and electronically controlled
- Operating mode: Man. / Vent.
- Safety valve: ≤8kpa
Ventilator Specification
- Patient type: Adult, pediatric
- Ventilation modes: IPPV, PLV, SIMV (flow-trigger), Manual
- Setting: Touch key & navigator wheel knob
- Tidal volume: 0-1500ml
- Ventilation frequency: 4-100bpm
- I:E ratio: 2:1-1:8
- Pmax: 5-70cmH O 2
- Ppeep: 0-20 cmH2O
- Inspiratory plateau: 0-60%Ti
- Trigger sensitivity: 2-30L/min
- High pressure: 21-80cmH2O
- Low pressure: 0-20cmH2O
- Monitoring: Display: 10.4” TFT color screen
Pressure: Paw, Ppeak, Pplat, Ppeep
Volumes: MV, Vt
Breath rate
Gas: FiO2
Compliance
Graph display: P-T, F-T, V-T waveforms&P-V,F-V loops
- Alarm: Pressure: Paw high/low limit
Volume: MV high/low limit,
High/Low limit of breath rate
Gas: FiO2 high/low limit
Low O2 supply pressure
Airway malfunction
Power failure
Vaporizer
- Support 2 vaporizer (Selectatec compatible )