Product Main

Specifications

HY-600I Door-Free Infusion Pump
Features
  • Precision Infusion Pump with unique door-free structure and big LCD screen, making it easier and more convenient to operate and in the meanwhile avoiding problems of pump stuck and fluid leakage.
  • Unique function of bedside infusion supervision, making patients and their relatives more assured while also reducing nurses’ workload. If used with SK-2000II and SK Data Collector Terminal, a bed No. Can be set in the machine and then its infusion progress and alarm information can be displayed on the SK-2000II located at Nurse’ s Station and SK Date Collector Terminal. The nurse, upon seeing and hearing the signal, can go and handle the unit accordingly.
  • All brands of standard IV sets are compatible. With only a simple calibration, it can guarantee an accurate, smooth and safe infusion.
  • When detecting air in line or occlusion, an audio and visual alarm signal shall be given instantly while at the same time the machine stops operation automatically, thus reducing the danger of air embolism to patients.
  • The infrared sensor shall give alarm signal when detecting excessive air pressure.
  • Application of double CPU, making the infusion process safer and more reliable.
  • Alarm signals are given in case of malfunction or operation error and the machine stops operation automatically.
 
Specifications
  • Applicable IV sets: All brands of standard IV sets
  • Mechanism: Peristaltic
  • Accuracy: +/-3%
  • Pressure: 39.2kpa~137.2kpa
  • Infusion rate: 1-600ml/h (increment 1ml/h)
  • Accumulated infusion volume: 1ml~9999ml
  • Volume limit: 1ml-9999ml (increment 1ml/h)
  • Displayed information: Rate, volume limit, accumulated volume, power indicator light, bed No, AIR, OCCL. OVER, EMPTY
  • Alarm functions: OVER, OCCL, Air, Empty, Low battery, control abnormal, no AC power supply, installation error, serious low battery
  • KVO rate: 1ml/h (keep vein open rate)
  • Bolus rate: 600ml/h
  • Weight: < 3.0kg
  • Dimension: 265*230*175mm(length*width*height)
  • Power supply: AC100V~240V, 50/60Hz
  • Fuse: T2AL250V~
  • Maximum power consumption: 25W, can run for 4h after being recharged
  • Battery: Rechargeable lithium polymer battery, 7.4V, 1600Ah
  • Battery recharge: When the pump is connected to the AC power supply, turn on the power and the battery will automatically recharge (about 8~14 hours to recharge fully. )
  • Electrical safety: Comply with the stipulation of IEC60601-1: 1996
  • Electromagnetic compatibility: Comply with the stipulation of IEC06061-1-2: 2006
  • Shell material: ABS plastic
  • Classification: Class II, Type BF
  • Operating conditions: Environment temperature: 5*C~40*C, air pressure: 86kpa~106kpa, relative humidity≤ 80%
  • Storage conditions: Environment temperature -40*C~55*C, air pressure 50kpa~106kpa, relative humidity≤ 95%
  • Applied standards: IEC60601-1: 2005

IEC60601-2-24: 1998

IEC60601-1-2:2006

ISO780-1997 ISO15223: 2000

ISO8536-4: 2004

ISO/IEC: Guide No. 37< Consumer Goods Usage Instructions