Product Main

Specifications

Infusion Pump (HY-600II)


Features of Infusion Pump
  • Three kinds of modes: Rate mode, Time Mode, Weight Mode.
  • Unique door free structure: Avoiding problems of pump stuck and fluid leakage.
  • Double CPU: Making the process of infusion safer and more reliable.
  • Accurate Infusion: Precise control of the infusion rate and infusion volume.
  • Flow raterange:1.0ml/h-2000.0ml/h
  • Driven by step motor: No pulsating wave will be educed even during low-rate infusion.
  • Alarm is given in the following situations: Infusion Completion, Occlusion, Low Battery, Air Bubble, Installation Error, Malfunction or Operation Error.
  • IV Set: All brands of standard IV sets are compatible
  • KVO rate can be selectable according to concrete requirements from 1.0ml/h to 5.0ml/h.
  • Wireless Module Control: When used with KS-2000II or SK date collector terminal, the infusion information will be transferred to the Infusion Supervision System in order to ensure the patients’ safety.
 
Specifications
  • Dimension: 120*40*190 mm (length*width*height)
  • Maximum weigh: <2.0kg
  • Displayed information: Rate, volume, E(accumulated infusion volume), Battery charge indicator, AC power indicator, OCCL. OVER, AIR, BED No.
  • FUSE: T 2AL 150V~
  • Maximum power consumption: 25VA, running more than 4hours after being charged.
  • Infusion rate range: 1.0-2000ml/h
  • Bolus: 1000.0ml/h
  • Operating conditions: Environment temperature 5℃~40℃, air pressure 86kpa~106kpa, relative Humidity≤80%
  • Electrical safety: Comply with EN60601-1 for medical equipment and EN60601-2-24
  • Applicable infusion sets: All standard IV sets
  • Accumulated infusion volume: 0.1~9999.0ml
  • Infusion accuracy: +/-3%
  • Alarm function: OVER, OCCL, Low battery, IV line installation error, Air, Empty, Machine malfunction
  • KVO rate:1.0ml/h-5.0ml/h
  • Selectable modes: Rate mode, Time mode, Weight mode.
  • Rate mode: 0.1~2000.0ml/h
  • Time mode: Volume limit: 0.1~999.9ml Time: 1~2000minutes
  • Body Weight mode: Drug limit: 0.1~999.9mg; Volume limit 0.1~999.9ml; Weight: 0.1~300.0kg; Dose: 0.1~9999.9; Selectable unit: mg/kg/h, ug/kg/min
  • Battery: Rechargeable lithium polymer battery, 7.4V, 1600Ah
  • Storage: Environment temp erature-20~55℃; Air pressure 50~106kpa; relative humidity≤95%