B-2000 Infant Incubator
STANDARD COMPOSITION:
●Servo Control Mode: air temp,skin temp;● Instant Parameter Displaying: set temp, air temp ,skin temp,heater power rate;● Audible & Visual Alarms: power failure, sensor failure, higher & lower temp, air& skin  temp override,fan failure,system failure;● >37℃ Preset Mode: power failure reset memory, alarm silence, keypad lock, self-test function;● Hood with dual slope: 6 hand-ports; left & right cabinets, top-shelf;● RS-232 Output Connector (Use for network system ). 

Specifications:
Power
Power requirements: AC220v 10%, 50Hz 2%
Power input: ≤ 450va
Refer to iec60601-1, this equipment belongs to class I, type b. Its operation is continuous.
Alarm
Power alarm
Power failure: Activates if primary power to the incubator
Fails or the power wire is accidentally
Disconnected from the wall receptacle.
Fan alarm
Fan failure: Activates if fan motor malfunctions. Sensor alarm
Air sensor failure: Air sensor malfunction, refer unit to service.
Sensor out of position: Activated when sensor is not in correct position to perform calibration or continue operation.
Skin sensor disconnected: Activated during skin control mode when skin sensor is disconnected from temperature controller. Skin sensor failure: Activated when skin sensor malfunctions.
Deviation alarm high air temperature: Activated if displayed air temperature is 1.5 centigrade above set point
Low air temperature: Activated if displayed air temperature 2.5centigrade below set point
Skin temperature: Activated if displayed skin temperature is 1.0 centigrade above set point
Low skin temperature: Activated if displayed skin temperature is 1.0over temperature alarm
Over air temperature: Activated when displayed air temperaturewas approach 38centigrade for set temperature ≤ 37centigrade, or approach 40centigrade for set temperature >37centigrade in air mode.
Over skin temperature: Activated when displayed skin temperature was approach 38centigrade for set temperature ≤ 37centigrade, or approach 39centigrade for set temperature >37centigrade in skin mode. System alarm
System failure: Activated if temperature control system malfunctions.
Temperature controlrangeand correlative specifications
Air temperature: 20centigrade ~37centigrade, for set temperature ≤ 37centigrade
37centigrade ~39centigrade, for set temperature >37centigrade
Skin temperature: 20centigrade ~37centigrade, for set temperature ≤ 37centigrade
37centigrade~38centigrade, for set temperature >37centigrade
Temperature rise time: ≤ 30min
Temperature variability: 0.5centigrade
Temperature uniformity: ≤ 1 centigrade
Skin sensor precision: ≤ 0.3centigrade
Environment requirement
Environment temperature
Operation temperature range+20centigrade ~+30centigrade
Store temperature range -40centigrade ~+70centigrade
Environment humidity
Operation humidity range: 30%~75%
Store humidity range: ≤ 80%
Atmospheric pressure range: 500hpa~1060hpa
Other specification
Noise level within hood environment ≤ 55dB(environment noise below 45dB carbon dioxide (CO2) level within the hood less than 0.5% when a mixture of CO2
In air is delivered at 750ml/min
At point 10cm above the center of the mattress.
Air velocit y over mattress less than: 0.35m/s
Infant bed max load: 10kg
Transfusion shelf max load: 2kg
Shelf max load:
Cabinet max load: 6kg