BNT-1000 Infant Radiant Warmer
STANDARD COMPOSITION
●Servo Control Mode:air temperature ● Instant Parameter Display: set temp, skin temp, heat power rate; ● Audible/Visual Alarms: temp deviation, sensor failure, power failure, temp override; ● Tilting Bed with Rotary Radiant Head ● X-ray Cassette Tray under infant bed, examination light of high illuminance; ● Use anti-blast quartz infrared tube as radiant heating source.
TECHNICAL SPECIFICATION:
This equipment belongs to ClassⅠ, Type BF, continuous operation common equipment.
POWER
Model BNT-1000-1E.........................................................................................AC 110 / 120 V, 50Hz / 60Hz
Model BNT-1000-2E......................................................................................... AC 220 / 240V, 50Hz / 60Hz
Input Power..................................................................................................................................... ≤1000VA
Model BNT-1000-1E.........................................................................................AC 110 / 120 V, 50Hz / 60Hz
Model BNT-1000-2E......................................................................................... AC 220 / 240V, 50Hz / 60Hz
Input Power..................................................................................................................................... ≤1000VA
TEMPERATURE CONTROL RANGE AND RELATIVE SPECIFICATION
Control mode....................................................................................................................................baby mode
Temperature control range.............................................................................................................25℃~37℃.
Skin temperature sensor measure range...........................................................................................20℃~42℃
Deviation between sensor measure temperature and control temperature.................................................≤0.5℃
Skin Sensor Precision........................................................................................................................... ≤0.3℃
Bed surface temperature Uniformity......................................................................................................... ≤1℃
Temperature Rise Time ....................................................................................................................... ≤45min
Bed tiltable.............................................................................................................continues adjustable in ±10°
Control mode....................................................................................................................................baby mode
Temperature control range.............................................................................................................25℃~37℃.
Skin temperature sensor measure range...........................................................................................20℃~42℃
Deviation between sensor measure temperature and control temperature.................................................≤0.5℃
Skin Sensor Precision........................................................................................................................... ≤0.3℃
Bed surface temperature Uniformity......................................................................................................... ≤1℃
Temperature Rise Time ....................................................................................................................... ≤45min
Bed tiltable.............................................................................................................continues adjustable in ±10°
ALARM
Sensor failure alarm.....................................................................alarm with audible and visual when sensor open
circuit, short circuit or disconnect, cut off power.
Over temperature alarm...................................................................alarm with audible and visual when displayed
temperature was approach 38℃.
Deviation alarm.....................................................................after temperature equilibrium, alarm with audible and
visual when displayed temperature was 1℃ above or
below control temperature, cut off power if above 1℃.
Power failure alarm..................................................................... alarm with audible and visual when power failure
ENVIRONMENT TEMPERATURE
(It is a suggestion that please do not use outside of the allow environment.)
Operate temperature range................................................................................................................18℃~30℃
Store temperature range.............................................................................................................. -20℃~+55℃
ENVIRONMENT HUMIDITY
Operate humidity range........................................................................................................... 30%RH~75%RH
Store humidity range............................................................................................................................ ≤93%RH
ATMOSPHERIC PRESSURE
Transport and Store atmospheric pressure range.......................................................................500hPa~1060hPa
Operate atmospheric pressure range.........................................................................................700hPa~1060hPa
Sensor failure alarm.....................................................................alarm with audible and visual when sensor open
circuit, short circuit or disconnect, cut off power.
Over temperature alarm...................................................................alarm with audible and visual when displayed
temperature was approach 38℃.
Deviation alarm.....................................................................after temperature equilibrium, alarm with audible and
visual when displayed temperature was 1℃ above or
below control temperature, cut off power if above 1℃.
Power failure alarm..................................................................... alarm with audible and visual when power failure
ENVIRONMENT TEMPERATURE
(It is a suggestion that please do not use outside of the allow environment.)
Operate temperature range................................................................................................................18℃~30℃
Store temperature range.............................................................................................................. -20℃~+55℃
ENVIRONMENT HUMIDITY
Operate humidity range........................................................................................................... 30%RH~75%RH
Store humidity range............................................................................................................................ ≤93%RH
ATMOSPHERIC PRESSURE
Transport and Store atmospheric pressure range.......................................................................500hPa~1060hPa
Operate atmospheric pressure range.........................................................................................700hPa~1060hPa
STREAM VELOCITY
Ambit environment stream velocity.......................................................................................................... ≤0.3m/s
Ambit environment stream velocity.......................................................................................................... ≤0.3m/s