Main Technical Specification
|
|
|
Testing item :
|
More than 200 items
|
|
Analysis method
|
End-point, fixed-time(two point rate) kinetics, non-linear,calibration,dual-wave length,double reagent,immune-turbidimetry, photo metric, sample reagent blank , numeration, absorbance
|
|
Wave length range
|
300-800nm
|
|
Optical system
|
Optical splitting system
|
|
Wave length
|
8 standard filters, 340, 380, 405,492, 510, 546, 578, 620nm
|
|
Wave length accuracy
|
±2nm, band width‹12nm
|
|
Flowcell
|
32ul quartz flowcell , optical path: 10mm
|
|
Suction amount
|
1-3000ul
|
|
Light source:
|
Tungsten-halogen lamp
|
|
Absorbency range
|
-0.214-3500Abs accuracy: external 0.001Abs, internal:0.001 Abs
|
|
Temperature
|
37,°C ,30°C,25°C(room), temperature accuracy:±0.1°C
|
|
Cross contamination
|
‹1%
|
|
Repeatability
|
‹1%
|
|
memory
|
>10000 results
|
|
Printer
|
Built in thermal printer
|
|
Control function
|
L-J, X, SD, CV%
|
|
interface
|
RS232
|
|
Power supply
|
220±22VAC, 50 ±1HZ 100W
|
|
Size :
|
380mm(L)*320mm(W)*150mm(H)
|
|
Working environment:
|
Temperature 0-40°C
|
|
Humidity
|
10-90%
|
|
Weight
|
8kg
|
- Weight: 8kg