Product Main

Specifications

LC98I Amino Acid Liquid Chromatography

 
P98I Binary Gradient Pump
1. SPECIFICATION
a. Operating pressure: 0-42MPa (6000 psi) maximum.
b. Flow rate range: 0.1-15ml/min,
c. Flow rate precision: 1%,
d. Flow rate accuracy: 25ml/368-387s(3%),
e. Compositional precision:0.5% absolute
f. Pressure solution: 0.1MPa,
g. Over pressure protection: pump shuts off if pressure exceeds limit specified by user-42.0 MPa(6000 psi) maximum.
2. Environment
a.Temperature requirements: ambient temperature range 10 to 35 degree
b. Operates in relative humidity from 5% to 80%;
c. Power requirements: 220Vac 5%,
d.Power frequency: 50 0.5Hz (voltage and frequency specified on the tag attached to the power cord)
3.  GENERAL PARAMETERS
a. Power Supply: 400W,
b. Dimensions: 450mm(D)×210mm(W)×390mm(H),
c. Weight: 20kg,
UV98I UV Detector
1. Specification
Wavelength range: 190-700nm
Spectral bandwidth:5nm
Wavelength accuracy: 2nm
Wavelength reproducibility: 0.2nm
Absorbance range:
Recorder output: 0.002au/mV to 2au/mV in steps of 1, 2, 5
Data system output: 2au/V, fixed
Noise: 5×10-5au (Peak-Peak)
Test Conditions:  Wavelength: 254nm
Mobile phase: methanol
Flow rate: 1ml/min
Time constants: 0.5s
Drift: 1×10-3au/hr
Test Conditions: Wavelength: 254nm
Mobile phase: methanol
Flow rate: 1ml/min
Time constants: 0.5s
Flow cell volume:4.5ul
Time constants: 0.05s, 0.5s, 5s
Autozero:Yes
2.  General Parameters
a. Power Supply: 200W,
b. Dimensions: 450mm(D)x180mm(W)×390mm(H),
c. Weight: 15kg
 
T98 Column Oven
Temperature controller and column oven combined in one or as individual unit.
a.  temperature range: ambient-150
b.  precision: ±0.5
c.  capacity of column oven: combined unit 300mm, long column 4 pieces.
 
N2000 Work Station
Double channel, Modem, GMP certificate.
 
Rheodyne 7725i Injector
Pressure up to 42MPa, nice feeling.