Sonoline C Fetal Doppler
Detail Description of Fetal Doppler:
Specification:
  • Anti-electroshock Type: Internally powered equipment.
  • Anti-electroshock Degree: Type B applied part
  • Harmful Liquid Proof Degree:
  • Main unit : Ordinary equipment (sealed equipment without liquid proof).
  • Probe: Prevent from water splashing, degree of protection: IPX4.
  • Degree of Safety in Presence of Flammable Gases: Equipment not suitable for use in presence of flammable gases.
  • Working System: Continuous running equipment.
  • EMC: Group I Class B.
  • Suitable Using Range: Suitable for use after the 12th week of pregnancy.
  • Physical Characteristic
  • Size: 135mm (Length) × 92mm (Width) × 29 (Height) mm
  • Weight: About 245g (including batteries)
 
Working Environment
  • Temperature: +5℃~+40℃
  • Humidity: ≤ 80%
  • Atmospheric Pressure: 70 kPa~l06kPa Transport and Storage:
  • Temperature: -10℃~+55℃
  • Humidity: ≤ 93%
  • Atmospheric Pressure: 50 kPa~l06kPa Display: 45mm× 25mm LCD display FHR Performance
  • FHR Measuring Range: 50~240BPM (BPM: beat per minute)
  • Resolution: 1BPM
  • Accuracy: ± 2BPM
  • Power consumption: < 0.8W
  • Auto Shut-OFF: After 1 minute no signal, power off automatically.
  • Battery Type Recommended: Two pieces of 1.5 V DC battery (SIZE AA LR6).
 
Probe:
  • Nominal Frequency: 2.0MHz
  • Working Frequency: 2.0MHz± 10%
  • P-: < 0.5MPa
  • Iob: < 10 mW/cm2
  • Ispta: < 50mW/cm2
  • Ultrasonic Output Intensity: Isata< 5 mW/cm2
  • Ultrasonic Output Power: P< 10 mW
  • Working Mode: Continuous wave Doppler
  • Effective Radiating Area of Transducer: 157mm2± 15%