HY-7700A Anaesthesia Machine
Detail Description
Detail Description
?Technical Specification
Physical & Environmental Specification of Anaesthesia Machine
- Dimensions(W×H×D): 1000×1400×770mm
- Weight: 148kg
- Storage temperature: -22~+55 ℃,
- Working temperature: +10~+40℃,
-
Storage humidity: ≤93%
Electronical Specification
- Mains: 100-240V AC, 50-60Hz
- Battery: 24V DC, minimum 120 min
-
Power consumption: 10A Max
Pneumatics Specification
- Gas supply: AIR, O2, N O: 0.28-0.6Mpa 2
- Flowmeters: AIR: 0-1L/min; 1-15L/min
- O2:0-1L/min: 1-10L/min
- N2O:0-1L/min: 1-12L/min
- Gas system: O2 deficiency alarm
- Hypoxic guard system
- O2 flush: 35-75L/min
- Working mode: Closed, Semi-closed, Semi-open
- Driven mode: Pneumatically driven and electronically controlled
- Operating mode: Man. / Vent.
-
Safety valve: ≤8.5kpa
Ventilator Specification
- Patient type: Adult, pediatric & neonate
- Ventilation modes: IPPV, PCV, SIMV (flow-trigger), PS, Manual
- Setting: Touch key & navigator wheel knob
- Tidal volume: 20-1500ml
- Ventilation frequency: 2-100bmp
- I:E ratio: 4:1-1:8
- Pmax: 5-70cmH2O
- Ppeep: 3-30 cmH2O
- Inspiratory plateau: off, 5%-60%Ti,
- Trigger sensitivity: 1-15L/min
- High pressure: 10-80cmH2O
- Low pressure: 0-70cmH2O
-
Monitoring: Display: 10.4” TFT color screen
Pressure: Paw, Ppeak, Pplat, Pmean,Volume: MV, VtBreath rateGas: FiO2ComplianceGraph display: P-T, F-T, V-T waveforms &P-V,F-V loops.Waveform frozen
-
Alarm: Pressure: Paw high/low limit
Volumes: MV high/low limit,High limit of breath rateGas: FiO2 high/low limitApneaAirway pressure highHigh PEEP levelPower failureLow O2 supply pressureVaporizer
- Support 2 vaporizer (Selectatec compatible )
- Agent type :Halothane, Enflurane, Isoflurane , Sevoflurane available.