HY-7700A Anaesthesia Machine
Detail Description
?Technical Specification
 
Physical & Environmental Specification of Anaesthesia Machine
       
  1. Dimensions(W×H×D): 1000×1400×770mm         
  2. Weight: 148kg         
  3. Storage temperature: -22~+55 ℃,         
  4. Working temperature: +10~+40℃,         
  5. Storage humidity: ≤93%         
    Electronical Specification         
  6. Mains:         100-240V AC, 50-60Hz         
  7. Battery: 24V DC, minimum 120 min         
  8. Power consumption: 10A Max         
    Pneumatics Specification         
  9. Gas supply: AIR, O2, N O: 0.28-0.6Mpa 2         
  10. Flowmeters: AIR: 0-1L/min; 1-15L/min        
  11.  O2:0-1L/min: 1-10L/min        
  12.  N2O:0-1L/min: 1-12L/min         
  13. Gas system: O2 deficiency alarm        
  14.  Hypoxic guard system        
  15.  O2 flush: 35-75L/min         
  16. Working mode:         Closed, Semi-closed, Semi-open         
  17. Driven mode: Pneumatically driven and electronically controlled         
  18. Operating mode: Man. / Vent.         
  19. Safety valve: ≤8.5kpa         
    Ventilator Specification         
  20. Patient type: Adult, pediatric & neonate         
  21. Ventilation modes: IPPV, PCV, SIMV (flow-trigger), PS, Manual         
  22. Setting: Touch key & navigator wheel knob         
  23. Tidal volume: 20-1500ml         
  24. Ventilation frequency: 2-100bmp         
  25. I:E ratio: 4:1-1:8         
  26. Pmax: 5-70cmH2O         
  27. Ppeep:         3-30 cmH2O         
  28. Inspiratory plateau: off, 5%-60%Ti,         
  29. Trigger sensitivity: 1-15L/min         
  30. High pressure: 10-80cmH2O         
  31. Low pressure: 0-70cmH2O         
  32. Monitoring: Display: 10.4” TFT color screen        
     Pressure: Paw, Ppeak, Pplat, Pmean,         
     Volume: MV, Vt        
     Breath rate        
     Gas: FiO2        
     Compliance        
     Graph display: P-T, F-T, V-T waveforms &P-V,F-V loops.Waveform frozen         
  33. Alarm:          Pressure: Paw high/low limit        
     Volumes: MV high/low limit,         
     High limit of breath rate        
     Gas: FiO2 high/low limit         
    Apnea         
    Airway pressure high        
     High PEEP level        
     Power failure        
     Low O2 supply pressure         
    Vaporizer                 
  34. Support 2 vaporizer (Selectatec compatible )         
  35. Agent type :Halothane, Enflurane, Isoflurane , Sevoflurane available.