Detail Description of Anaesthesia Machine:  
Features
 
  • The first Chinese anesthesia workstation with CE mark meeting American FDA standard. FDA filing in process.
 
  • PCV and SIMV+PS modes provide more flexibility to ventilation strategy, suitable for neonate, pediatric and adult operation.
 
  • Fresh gas, altitude and compliance compensation functions guarantee an accurate delivery of tidal volume
 
  • Auxiliary oxygen inhalation device supports oxygen through entire operation for general, local and spinal anesthesia process.
  • “ Zero Maintenance” AGSS(optional) and closed-refill style vaporizer(optional) provide more safety protection
 
  • Constant temperature maintaining function in the CO2 absorption system avoids water vapor coagulation and increases patient comfort
  • Integrated stainless steel table top prevents drug erosion and scratch by hard objects
  • Optional selection of O2, CO2 and anesthesia gas monitoring function
 
  • Integrated various devices related to in-operation stage provides electric power and gas power source
  • GCX connection support for easy upgrade to anesthesia workstation
 
 
 
Technical Specification
 
  • Physical & Environmental Specification
  • Dimensions (WxHxD): 650x1380x690mm
 
  • Weight:                           148kg
  • Storage temperature:   -10~+60℃
 
  • Working temperature:   +10~+40℃
 
  • Storage humidity: 15%~90%
 
  • Electrical Specification
 
  • Mains: 100-240V AC, 50/60Hz
 
  • Battery: 24V DC, minimum120min
 
  • Power consumption: ≤ 70W
 
  • Pneumatic Specification
 
  • Gas supply: O2, N2O, AIR: 0.28-0.6Mpa
 
  • Flowmeters:   O2: 0-1L/min; 1-10l/min
 
    N2O: 0-1L/min; 1-10L/min
 
      AIR: 0-1L/min; 1-10L/min                                       
 
  • Gas system:     O2 deficiency alarm
 
                                            Hypoxic guard system
 
                                            O2flush: 25-75L/min
 
  • Working mode: Closed, Semi-closed, Semi-open
 
  • Driven mode: Pneumatically driven and electronically controlled
 
  • Operating mode: Man. / Vent.
 
  • Safety valve: ≤ 8kpa
 
  • Ventilator Specification
  • Patient type: Adult, pediatric & neonate
 
  • Ventilation modes: IPPV, PCV, SIMV(flow-trigger), PS, Manual
  • Setting: Touch key & navigator wheel knob
 
  • Tidal volume: 20-1400ml
 
  • Ventilation frequency: 4-60bmp
 
  • I: E ratio: 4:1-1:5
  • Pressure range: 5-70cmH2O
 
  • Ppeep: off, 3-30cmH2O
 
  • Inspiratory plateau: off, 5%-60%Ti
 
  • Trigger sensitivity: 1-15L/min
 
  • High pressure: 10-80cmH2O
 
  • Low pressure: 0-70cmH2O
 
  • Monitoring: Display: 10.4” TFT color screen
 
                                  Pressure: Ppeak, Pplat, Pmean, PEEP
 
                                  Volume: MV, Vt
 
                                  Breath rate
 
                                    Gas: FiO2
 
                                    Graph display: P-T, F-T, V-T waveforms & P-V, F-V loops. Waveform frozen
 
  • Alarm:         Pressure: Paw high/low limit
 
                            Volumes: MV high/low limit
 
                            High limit of breath rate
 
                            Gas: FiO2 high/low limit
 
                            Apnea
 
                            Airway pressure high
 
                            High PEEP level
 
                              Power failure
 
                                Low O2 supply pressure
 
  • Vaporizer
  • Support 2 vaporizer (Selectatec compatible)
  • Agent type: Halothane, Enflurane, Isoflurane, Sevoflurane available.