Product Main

Specifications

Feaures of Medical Ventilator:
 
  • The first ICU medical ventilator with CE mark in China
  • Pneumatically driven and electronically controlled for child and adult
  • Detachable 10.4”TFT color screen decreases cross-infection effectively
  • Easy setting by navigator wheel knob & touch key
  • Three-level light and sound alarm, with concise words describing problems
  • Sensitive pressure trigger and flow trigger with base flow
  • Lung mechanics monitoring and loops
  • Emergency start function
  • Synchronized nebulizer
  • Built-in inspiration and expiration flow sensor, durable and water coagulation resistant
  • Built-in battery
 
Technical Specification
 
Modes of Ventilation
  • Volume Controlled (VCV), Assist/Control
  • Pressure Controlled (PCV) , Assist/Control
  • Synchronized Intermittent Mandatory Ventilation (SIMV)
  • Volume-controlled breaths (V-SIMV)
  • Pressure- controlled breaths (P-SIMV)
  • Pressure Support (PSV)
  • Spontaneous ventilation (SPONT)
  • Continuous Positive Airway Pressure (CPAP)
  • Combined settings: VCV + SIGH, SIMV (VCV)+PSV, SIMV (PCV)+PSV
  • Sigh
 
Parameter
  • Tidal volume:50~1500ml
  • Respiratory rate:4~100bpm(used in VCV & PCV)
    • 1~40bpm (used in V-SIMV & P-SIMV)
  • Inspiration time:    0.1~12s
  • Inspiratory pause time:0~4s
  • FiO2:21%~100%
  • Trigger sensitivity: Pressure (-2~0kPa, above PEEP)
    • Flow (0.5~20LPM, with base flow)
  • PEEP:0~4kPa
  • Psupport:0~6kPa
  • Pcontrol:0.5~6kPa
  • Nebulizer:30 minutes
  • Emergency start: Yes
  • Key lock: Yes
 
Monitoring
  • Pressure values: Paw, Pplat, Pmean, PEEP
  • Volume/Flow values: VTI, VTE, MV, MVspont
  • Time values: Ftotal, Fspont, I:E
  • Oxygen monitoring: without O2 sensor
  • Real time curves: Pressure - Time, Flow - Time, Volume -Time, Pressure - Volume loop, Flow - Volume loop
  • Respiratory mechanics: dynamic and static compliance, resistance, auto PEEP
 
Alarm
  • MV high, MV low, Circuit disconnected, Paw high, Paw low, Airway pressure continue high, VTE high, Fspont high, Tapnea, FiO2 low, FiO2 high, Power supply failure, Nebulizer on, Battery low, Battery exhausted, Air supply down, O2 supply down, Air & O2 supply down
 
Technical data
 
  • Screen:10.4”TFT color screen (detachable)
  • Gas supply:O2 ,Air (All gas must be medical level) , 0.28~0.6MPa
  • Power supply:AC110~240V, 50Hz/60Hz, 65VA
  • Maximum security pressure: 8kPa
  • Compliance:4mL/100Pa
  • Noise: ≤65dB(A)
  • Communication interfaces:RS232 port, VGA port
  • Dimensions(H×W×D):400×303×250mm
  • Weight:15kg
 
Environment requirement
Temperature:
5℃~40℃(Operation)
 
-10℃~55℃ (Storage)
Relative humidity:
≤80%, non-condensing(Operation)
 
≤93%, non-condensing (Storage)
Atmospheric Pressure:
86~106kPa(Operation)
 
86~106kPa(Storage)
Altitude    Operation:
500~800mmHg / 3565~ -440m (Operation)
 
375~800mmHg / 5860~ -440m (Storage)