Anaesthesia Machine Features:
The first Chinese anesthesia workstation with CE mark meeting American FDA standard. FDA filing in process.
PCV and SIMV+PS modes provide more flexibility to ventilation strategy, suitable for neonate, pediatric and adult operation.
Fresh gas, altitude and compliance compensation functions guarantee an accurate delivery of tidal volume
Auxiliary oxygen inhalation device supports oxygen through entire operation for general, local and spinal anesthesia process.
“Zero Maintenance” AGSS(optional) and closed-refill style vaporizer(optional) provide more safety protection
Constant temperature maintaining function in the CO2 absorption system avoids water vapor coagulation and increases patient comfort
Integrated stainless steel table top prevents drug erosion and scratch by hard objects
Optional selection of O2,CO2 and anesthesia gas monitoring function
Integrated various devices related to in-operation stage provides electric power and gas power source
GCX connection support for easy upgrade to anesthesia workstation
Technical Specification
Physical & Environmental Specification
Dimensions (WxHxD): 650x1380x690mm
Weight: 148kg
Storage temperature: -10~+60℃
Working temperature: +10~+40℃
Storage humidity: 15%~90%
Electrical Specification
Mains: 100-240V AC, 50/60Hz
Battery: 24V DC, minimum120min
Power consumption: ≤70W
Pneumatic Specification
Gas supply: O2, N2O, AIR: 0.28-0.6Mpa
Flowmeters: O2: 0-1L/min; 1-10l/min
N2O:0-1L/min; 1-10L/min
AIR:0-1L/min; 1-10L/min
-
Gas system: O2 deficiency alarm
Hypoxic guard system
O2flush: 25-75L/min
Working mode: Closed, Semi-closed, Semi-open
Driven mode: Pneumatically driven and electronically controlled
Operating mode: Man. / Vent.
Safety valve: ≤8kpa
Ventilator Specification
Patient type: Adult, pediatric & neonate
Ventilation modes: IPPV, PCV, SIMV(flow-trigger), PS, Manual
Setting: Touch key & navigator wheel knob
Tidal volume: 20-1400ml
Ventilation frequency: 4-60bmp
I:E ratio: 4:1-1:5
Pressure range: 5-70cmH2O
Ppeep: off, 3-30cmH2O
Inspiratory plateau: off, 5%-60%Ti
Trigger sensitivity: 1-15L/min
High pressure: 10-80cmH2O
Low pressure: 0-70cmH2O
Monitoring: Display: 10.4”TFT color screen
Pressure: Ppeak, Pplat, Pmean, PEEP
Volume: MV, Vt
Breath rate
Gas: FiO2
Graph display: P-T, F-T, V-T waveforms & P-V, F-V loops. Waveform frozen
-
Alarm: Pressure: Paw high/low limit
-
Volumes: MV high/low limit
High limit of breath rate
Gas: FiO2 high/low limit
Apnea
Airway pressure high
High PEEP level
Power failure
Low O2 supply pressure
-
Vaporizer
-
Support 2 vaporizer (Selectatec compatible)
-
Agent type: Halothane, Enflurane, Isoflurane, Sevoflurane available.