Syringe Pump Features
- Double channel (multi-channel), sub-rate injection.
- Four kinds of injection modes: rate mode, time mode, body weight mode, link mode.
- Innovative design of outer appearance gives a strong sense of science and technology.
- Double CPU makes the process of injection safer and more reliable
- The use of TFT color LCD makes the screen display more remarkable
- Digital menu, easy operation, functions fully meet the clinical demand
- Integrative pump device makes more smooth operative and easier installation and maintenance.
- Brand-new STM32 ARM-cortex M3 platform, optimized circuit design, operation rate more faster and stable.
- Ample and user-friendly indicator information, accurate and complete alarm information to ensure safe injection.
- Unique wireless injection supervision function, achieving accurate injection and synchronization injection supervision to patients.
Specifications
- Max injection rate: 1500ml/h (different injectors have different maximum rate)
- Accumulated volume: 0.1ml~999.9ml
- Max size of outer shell: 320mm*200mm*110mm (length*width*height)
- Applicable injectors: 5ml, 10ml, 20ml,30ml, 50ml
- Injection accuracy:+/-3%( up to +/-1% after calibration)
- Injection increment: 0.1ml/h
- Volume limit: 0.1ml~999.9ml
- Classification: class II, type BF
- Battery: AC100~240V, 50/60Hz
- Battery recharge: rechargeable lithium polymer battery, 7.4V, 1600mA/h
- Max power consumption: single-channel works for continuous 5 hours at the rate of 25ml/h after fully recharged; double-channel works for continuous 3 hours at the rate of 25ml/h after fully recharged.
- Range of flow rate: 5ml injector: 0.1ml/h~100ml/h
10ml injector: 0.1ml/h~200ml/h
20ml injector: 0.1ml/h~400ml/h
30ml injector: 0.1ml/h~600ml/h
50ml injector: 0.1ml/h~1500ml/h
- Bolus rate: 5ml injector: 100ml/h
10ml injector: 200ml/h
20ml injector: 400ml/h
30ml injector: 400~600ml/h (adjustable)
50ml injector: 400~1500ml/h (adjustable)
- Alarm information: near completion, completion, OCCL, injector off, low voltage, No AC power supply, communication error, link abnormal, pump stuck, reading error of the parameters, etc.
- Mode selection: rate mode, time mode, body weight mode, link mode
- Rate mode: rate: 0.1ml/h~1500ml/h (max value depends on injector specification)
- Time mode: volume: 0.1ml~999.9ml; time: 1min~2000min
- Body weight mode: weight: 0.1kg~300.0kg; drug: 0.1mg~999.9mg; volume: 0.1ml~999.9ml; dose: 0.1~1500.0(restricted to unit, drug, solution, weight and injector); unit: mg/kg/h, ug/kg/min
- OCCL value: 40kpa~160Kpa, display the voltage status and the pressure value
- KVO rate: adjustable range: 0.1ml/h~5.0ml/h ( adjustable: 0.1ml/h per each segment, press "0" to cancel KVO function)