Product Introduction:
The LH600 CO2 sensor is the ideal low sampling-rate capnography solution for your patient monitoring system. LH600 can show an accurate capnogram with all respiration rates up to 120 breaths per minute. Cutting-edge technology enables LH600 to provide accurate measurement of instantaneous CO2 concentration, respiration rate, end-tidal CO2, inspired CO2, inspiration time and expiration time. With a sampling-rate low to 50 mL/min, LH600 is ideal for use on adult, pediatric and infant patients; patients undergoing procedural sedation; and short- term monitoring of intubated patients.
Product Features:
1.State-of-the-art Non-dispersive infrared (NDIR) technology, dual wavelength, no moving parts
2.Breath-by-breath CO2 monitoring
3.Sample rate low to 50 mL/min
4.Barometer pressure and temperature compensation
5.Range calibration in 0 ? 15% by manufacturer
6.Fast warmup time ensures full accuracy within 5 minutes
7.Oxygen, N2O and anesthetic compensation for accurate measurement
8.Low power consumption less than 2W
9.Fully compatible with Respronics Loflo sensor and CPT module
Technical Specifications:
|
Transducer Type
|
LH600 Sidestream CO2 sensor
|
|
Sample Rate
|
50 mL/min, ?10mL/min.
|
|
Principle of Operation
|
Non-dispersive infrared (NDIR) single beam optics, dual wavelength, no moving parts
|
|
Initialization Time
|
Capnogram displayed in less than 5 seconds
Reaching full specifications within 5 minutes.
|
|
CO2 Measure Range
|
0 - 114 mmHg
0 - 15%
0 - 15.2kPa
|
|
CO2 Resolution
|
0.1 mm Hg 0 to 50 mm Hg
0.25 mm Hg 50 to 114 mm Hg
|
|
CO2 Accuracy
|
0 ? 40 mmHg ?2 mmHg
40 ? 114 mmHg ?5% of reading
|
|
Respiratory Rate
|
Range 2 to 120 Breaths Per Minute (BPM)
Accuracy ?1 breath
|
|
Calibration
|
No routine user calibration required
|
|
Compensation
|
Automatic Barometer Pressure 400 - 800 mmHg.
User selectable O2 / N2O compensation.
|
|
Voltage Requirements
|
5.0 VDC ?5%
|
|
Power Consumption
|
Less than 1.5 Watts typical.
|
|
Temperature and Humidity
|
Operating: 0º to 40?, 10 to 90% RH, non-condensing
Storage: -40º to 70ºC, <90% RH, non-condensing
|
|
Interconnection
|
Pin out:
1. TxD Serial data from LH600
2. RxD Serial data from Host
3. AGND Analog return
4. DGND Digital return
5. VA 5.0V
|
|
Data Interface
|
TTL level, bi-directional, 19200 baud rate, standard N-8-1.
Fully compatible with Respironics LoFlo communication protocol
|
|
Data Output
|
CO2 gas concentration (mm Hg), End-tidal CO2, Inspired CO2, Respiratory Rate, barometric pressure compensation. Compensation for O2 (0 - 100%) and N2O (On / Off) when supplied by host.
|
|