Product Main

Specifications

Features
ETCO2
Insp. CO2
RR
SPO2
PR
Technical Specifications
 
Pulse Oximetry
SPO2 Range: 1% to 99%
Resolution: 1%
Accuracy: 70% to 99% range: ?2%; 
 1% to 69% range: ?5%;
Method: Dual wavelength LED
 
NIBP (Non-invasive Blood Pressure)
Technique: Oscillometric during inflation
Range: Adult/Pediatric: 40 to 240 mmHg
 Neonate: 10 to 150 mmHg
Measurement Cycle: < 40 sec. typical
Automatic Measurement
Cycles (Selectable): 2, 3, 5, 10, 15, 30 min; 1, 2, 4 hr
STAT Mode: 5 minutes of continuous readings
Max. Allowable Cuff 
Pressure: Adult: 300mmHg
 Pediatric: 240 mmHg; 
 Neonate: 150 mmHg
Resolution: 1 mmHg
Transducer Accuracy: ?2 mmHg or ?2% of reading, 
 
Heart (Pulse) Rate
Range: 30 to 250 bpm 
Accuracy: ?2 bpm or ?3%, whichever is greater
 
Capnography (CO2) 
Units: mmHg; kPa; %
Method: Non-dispersive infrared; auto calibrating
N2O Compensation: Manual; On/Off, user selectable
Range: 0 to 114 mmHg; 0 to 15%
Flow, Sample Rate: Sidestream; 50 ml/min
Accuracy: ?2 mmHg or 5% of reading
 
Respiration Rate
Rate: 2 to 120 bpm (CO2)
Resolution: 1 breath/min
 
Trends
Memory: 96 hours with a NIBP reading
 every 5 minutes
Display: tabular, graphical
 
Interface & Display
Keys: 4; membrane activated
Rotary Knob: Push and rotate; 24 steps/turn
Screen: 7.0 inch active color TFT
Resolution: 640 x 480 pixels
Waveforms: 3, maximum
Waveform Types: CO2; Pleth; 
Sweep Speed: 6.25, 12.5, 25 mm/s
Languages: English
 
Printer (optional)
Type: Thermal line printer
Paper Speed: 6.25, 12.5, 25 mm/sec
 
Physical Properties
Weight: 1.7 kg
Size: 17.5 cm (H) x 23 cm (W) x 11 cm (D)
 
Power Requirements
Voltage: 110/220 VAC; 50 Hz
Power Consumption: 8 Watts, typical
Battery: 1 sealed li-battery
Battery Life: 4 hours, typical
Recharge Time: 4 hours, typical