Product Main

Specifications

R Type Thermocouple Temperature Sensor
Type R thermocouple is the PtRh13-Pt thermocouple, it also used as a temperature transmitter, in general, type R thermocouple is connected with the display instrument, record instrument and computer by RC compensation cable. It can be directly measured temperature of the gas, liquid , melting liquid and solid surface in various production process. Now the type R thermocouple have three forms: assembly thermocouple, sheathed thermocouple , assembly and sheathed combined thermocouple.
 
Main technology index
 
Temperature range and tolerance
Name
Calibration
Tolerance
 
PtRh10-Pt
WRP or WRPK
 
 
R
Class l
Class ll
Tolerance
Temperature range° C
Tolerance
Temperature range° C
± 1° C or ± (1+(t-1100)× 0.003)
0~1400
± 1.5° C or ± 0.25%|t|
0~1400
 
Thermal response time ( type R assembly thermocouple)
 
Diameter of protection tube mm
Material of protection tube
Thermal response time S
16
Non-metal
≤ 240
25
Non-metal
≤ 300
 
 
Thermal response time ( type R sheathed thermocouple)
Thermal response time t0.5S
Thermowell Diameter(mm)
 
Shell-connecting type
 
Insulation type
3.0
0.6
2.0
4.0
0.8
2.5
5.0
1.2
4.0
6.0
2.0
6.0
8.0
4.0
8.0
 
 
Max temperature of type R thermoeouple wire
Thermocouple wire diameter mm
Recommended Max temperature ° C
0.5
1500
 
1) Strong junction box to enclose parts that might produce sparks,
electric arc, and dangerous high temperatures
2) When explosions within the box occur, fire and high temperatures
caused by explosion are extinguished and cooled within the gap
3) Various explosion-proof types are available
4) Good explosion proof performance
5) Spring thermal element with good shockproof performance
6) Wide measuring range