Product Main

Specifications

High Frequency leakage
 Bipolar : Less then 60mA
 Monopolar : Less than 150mA
Low frequency leakage
Source current ,patient leads, all out puts tied together
Normal polarity ,intact chassis ground<10μA
Normal polarity ,ground open<50μA
reverse polarity, ground open<50μA
skin current,140Vaplied,all inputs<50μA