Description
ACD-2C Storage Type Digital Pressure Gauge adopts the storage module built-in method on the basis of the ACD-200 precise digital pressure gauge, the gauge can store 10000 data items. The stored data can be transmitted to the computer through the communication interface, and then the data analysis and report processing shall be conducted by the matched software.
Applications
l Wellhead pressure data acquisition, storage and analysis of oil well and gas well
l Pressure data acquisition, storage and analysis of city tap water and natural gas pipeline network
l Laboratory data communication pressure acquisition, data storage and processing can be realized on computer
l Calibration standard instrument of all industries.
Features
l Magnetic pen induction button design
l Visually display the percentage bargraph
l Acquisition speed can be set
l Always online communication technology, bus support 255 devices with RS485
l Super power management technology, service life of battery can be 4 to 6 years
l Optional rechargeable battery, fully charged battery can be used for 2 months
l Signal full isolation technology
l Anti electromagnetic interference and radio-frequency interference technology
l Zero point homeostasis technology
l Temperature compensation technology
Data Sheet
1. Measurement Range: Gauge Pressure -100Kpa~0~1kPa~260MPa( Optional within the range)
Differential Pressure: 0~500Pa~3.5MPa
Absolute Pressure: 0~10kPa~60MPa
2. Communication Method: RS485, RS232, USB and matched communication software
3. Accuracy Grade: 0.05, 0.1, 0.2, 0.5
4. Acquisition Speed: High-speed type: 25ms~100s/time( can be set as multiple of 25ms
Normal-speed type:1s~7200s/time( can be set as multiple of 1s)
5. Overload Pressure: 1.5~3 multiple span, which depends on measurement range.
6. Display Mode: dynamic 5 digits display, percentage bargraph of pressure;
7. Built-in Power Supply: rechargeable lithium battery or high power lithium battery
8. Medium Temperture:-40℃~85℃
9. Working Conditions: when choose rechargeable battery, temperature:-10℃~50℃; Humidity:<90%
When choose industrial lithium battery:-30℃~60℃; Humidity:<90%
10. Explosive-Proof Grade:ExiallCT6
11. Process Link:M20*1.5(external thread) or in accordance with customers demands
12. Optional Accessories: communication software, aluminum alloy suitcase, etc.