Product Main

Specifications

System function
 
Throughput: up to 225 tests/Hour(not including ISE tests)
Test Method: End point, Fixed time, Kinetic
Test Sequence: Discrete, Random access, STAT can be inserted at any time. Test sequence can be sorted by samples or items
Spectrometer: enclosed array of spectrometer and photodiodes, maintenance free.
Cuvette washing: auto wash the cuvettes through 8 segments washing pipe and continuous use them(optional)
Reaction detect: monitor 81 cuvettes’s reaction at the same time
Photometric process: the interval time between two consecutive photometric points are exactly the same.
Pipeline system: pressure auto released pumps and double filters to protect pumps and valves
Gearing: plastic gear leads lower noise and accuracy of positioning of moving parts
Environmental resistance: resistant form dust and water vapor
 
Sample/Reagent/Stirrer system
 
Sample volume:2μL~30μL, step by 0.1μL
Sample disc: 60 positions for primary tubes and secondary cups
Reagent volume:20μL~300μL, step by 1μL
Reagent disc: provide 60 reagent bottle positions
Reagent bottle: particular reagent bottle without dead volume
Reagent Cooling: continuous cooling system according to the principle of peltier
Probe: inner and outer walls are highly polished leads carry-over<0.1%, level detection and vertical and horizontal collision protection
Stirrer: paddle stirrer coated with Teflon, stir immediately after the sample was added in
Water consume: <1.5L/hour
Carry over: Optimized item test sequence and reinforced washing process, lower the carry over and avoid probe being clotted
 
Reaction system
 
Cuvette: permanent UV free glass cuvette
Reaction disc: Discal reaction disc with 81 disposable cuvettes
Reaction volume:150μL~600μL
Heater: solid and sealed heater leads temperature raise rapidly and steady and maintenance free
Temperature: 37℃±0.1℃
 
Optical system
 
Light source: 12V 20W, halogen lamp with lifetime more than 2000 hours
Spectrometer: post spectrometer with arrayed and sealed filters, maintenance free
Wavelength: 340nm, 405nm, 450nm, 510nm, 546nm, 578nm, 630nm, 670nm
Absorbance range: 0~6
 
Calibration/QC
 
Calibration type: single point linear, two points linear, multi-points linear, Logit-4P, Logit-5P, Spline, Exponential 5P, Polynomial 5P, auto choose the best fittest calibration curve
QC type: real time QC, daily QC, day to day QC
QC chart: L-J, Cumulative sum check, Twin Plot
QC rule: Customized QC rule in order to adapt different item’s requirements
 
Operational system
 
Operational system: VISTA, Windows XP, Windows2000
Memory: auto save and recover the results
Report format: Customized report format
Status Check: real time check the working status of the analyzer
 
Power/ Dimension
 
Voltage: AC 85V~264V, 55Hz±8Hz, Auto adapt according to the input voltage
Power Consume: ≤350VA
Dimension: 850mm×650mm×1067mm (L×W×H)
Net weight: 90Kg