Instrument introduction
EDX320L incorporates Delptek's years of experience in RoHS testing with unique product configurations and fully functional software. Its ergonomically design and friendly software interfaces grant the users a simple measurement job.
Performance characteristics
Professional RoHS detection and also applicable to Plating Thickness and full-element analysis
In-built SNE improves the signal processing ability up to 25 times
Automatic collimators and filters switch for different samples
Electro-cooling Si PIN semiconductor detector instead of liquid nitrogen cooling detector
Intelligent RoHS analysis software matches with the hardware well
Technical specifications
Measurable elements: 75 elements from Sulfur to Uranium
Analysis range: 1ppm-99.99%
Detection limit of hazardous substances (Cd/Pb/Cr/Hg/Br) in RoHS directive reaches 1ppm.
Measurement time: 60-200s
Energy revolution: 140±5eV
Tube voltage: 10-50 kV
Tube current: 50-1000μA
Standard configurations
Signal-to-Noise Enhancer (SNE)
Unique light path enhancement system
Electro-cooling Si PIN semiconductor detector
Inbuilt high-resolution camera
Automatic collimators and filters switch
Precise movable platform
Enhanced metal element sensitivity analyzer
Independent matrix effect correction models
Arbitrary optional analysis and identification models
Application fields
Hazardous substances detection in electric and electronic fields, plating industries, various materials(gold), plastics, glass,  and others
Plating industry