Features
● Taking IPC to drive concave grating, it has tunable wavelength, small error, high optical performance and long life.
● It supports full Chinese operations, comprehensive report in Chinese, manual entering of other measurements and their printing. It has built-in thermal printer and has optional multiple external printers.
● It has complete self-test function, and can carry out fault detection for its all systems after booting. It takes automatic analytic process, and can automatically alarm for the abnormal conditions (such as bubbles, substrate depletion, etc.).
● The large true color LCD screen can display characters and real-time response curve simultaneously, which are more convenient to operate. 8-17 inch LCD are optional.
● It has reasonable software designation, the function of patients historical data inquiry and auto-contrast. It can automatically review whether the test items are consistent with the applications or not to eliminate missing test.
It stores the results of at least 100 thousand, including quality control and standard curve, can count the workload and test income of anytime, and can print LJ quality control chart. Self-editing test items can be arbitrarily extended.
● Reagents are completely open, import, domestic and self-made reagents can be used.● With its power-down protection, the parameters and results will not be lost under the situation of power failure.
● With air automatic isolation, it can avoid cross-contamination.
Specifications
● Test methods: endpoint method (including absorbance, linearity density, nonlinearity density, can take dual-wavelength measurements), fixation time method (two points method), continuous monitoring method (dynamic method).
● Source: 20W/12V halogen lamps, with life larger than 3000 hours.
● Optical Systems: 330-800nm, the wavelength accuracy of ± 2nm.
● Absorbance range: 0-2 .500 Abs, resolution 0.001Abs.
●Shade selection pool: Titanium / quartz colorimetric cylinder, 1.5mm light path, 32ul.
● Test methods: endpoint method (including absorbance, linearity density, nonlinearity density, can take dual-wavelength measurements), fixation time method (two points method), continuous monitoring method (dynamic method).
● Source: 20W/12V halogen lamps, with life larger than 3000 hours.
● Optical Systems: 330-800nm, the wavelength accuracy of ± 2nm.
● Absorbance range: 0-2 .500 Abs, resolution 0.001Abs.
●Shade selection pool: Titanium / quartz colorimetric cylinder, 1.5mm light path, 32ul.
● Temperature Control System: using PELTIER component to control temperature 25 ℃, 30 ℃, 37 ℃ ± 0.5 ℃.
● Measuring accuracy: error ≤ 3%, drift <2mAbs / h
● With built-in 56 mm thermal printer, it can print parameters, measurement curve, calibration curve and the comprehensive report.
● Ambient temperature for working:15 - 30 degrees C Humidity: 20% - 85%
● Measuring accuracy: error ≤ 3%, drift <2mAbs / h
● With built-in 56 mm thermal printer, it can print parameters, measurement curve, calibration curve and the comprehensive report.
● Ambient temperature for working:15 - 30 degrees C Humidity: 20% - 85%
● Power: 220v ± 15%, 50Hz, 70VA.