This product is the use of the motor drives the pump to generate air flow and pressure, the atomizing tank liquid atomization, produce fine particles ejected drops, again through the breathing of the patient to the lesion site, such drugs and lesion body directly contacts, to work, to treat diseases of the respiratory system.

No. project

1 power supply ( different power supply using different specifications of motor ) 100/60HZ 120V/60HZ

230V/50HZ 220V/60HZ
2 the power consumption of 180 VA MAX
3 spray tank model Westmed ( Lavender ) OR Med2000 ( deep purple )
4 working pressure range 10 to 16 Psi ( 0.07 to 0.12 Mpa )
5 the maximum pressure range 29 to 43 Psi ( 0.20 to 0.30 Mpa )
6 work flow range 4.5 to 7.0Lpm
7 atomization efficiency of ≥ 0.2 ml/min ( 2% saline )
The 8 noise≤ 60 dB ( 1 meters )
10 to 90% RH 9 working humidity
10 temperature 10 ℃ to and 40℃( 50 ℉to 104℉)
10 to 95% RH 11 storage humidity
The 12 storage temperature of -25℃ to℃( 70℉to -13 158℉)
13 dimensions ( L ×W ×H ) of 188× 106× 188 mm ( 7.4× 4.2 × 7.4in )
The weight of 14 1.5 kg ( 3.3 lbs )
15 life for more than five years

16 safety UL1431, CE, FDA medical

Operating instructions:
A. charging and atomizing tank:

1 open the mechanical respirator attached storage bag, out of which the trachea and aerosol cans;

2 of the trachea is inserted into the mechanical respirator front air outlet, at the same time the trachea in mechanical respirator handle pipe groove;

3 pipe and another end is inserted into the atomizing tank;
B. atomized can join potion:
1 undo spray tank;
2 the amount of liquid into the atomizing tank;

3 an injection atomizing tank liquid atomizing tank top shall not exceed the scale line, otherwise will reduce the atomizing rate;

4 cover spray tank;
5 the atomizing tank on mechanical ventilator handle tracheal card slot;
C. boot:
1 is inserted into the plug power supply;

2 the power switch with " I " or " ON" end is pressed (power switch on the lights ), the machine begins to work;

3 removed the atomizing tank started treatment;
D. shutdown:

1 when the treatment is completed, the power switch is printed with " O " or " OFF" end is pressed (power switch on the lamp ), the machine to stop working;

2 if the treatment of unfinished drops have after use, must first down and then joined the potion after continued use;

The 3 machines at once the most long time should not exceed half an hour;

E. cleaning aerosol cans:

1 open atomizing tank, the tank end after atomization atomization medicine cleared;

2 if necessary, use a neutral detergent or disinfectant spray cans on a thorough cleaning.