Product Main

Specifications

Main Property
1. A variety of workflows- TPD, TPR, TPO, TPSR, Pulse Titration, Pretreatment and BET Specific Surface Area
2. Multiple Independent Gassource- In addition to carrier gas and reaction gas, each for independent control and mixes in scale.
3. Multiple Gas Joints - 3 for carrier gas and 3 for reaction gas
4. Steam Generator – It is equipped with a saturated steam generator and convenient to use reaction gas like liquid NH3.
5. Can be connected to GC or MS – It can be connected to the gas chromatography (GC) or mass spectroscopy (MS).
6. Auxiliary Detector – The instrument can output a voltage signal which is convenient to connect gas auxiliary detector, such as FID.
7. Material Structure - Stainless steel piping, seal for fluorine rubber, ask company for special custom if applied in special environment.
 
Precise Control
1. Flow Control – Using mass flow controller (MFC) software control, the flow is adjustable between 10 to 100sccm and software controls gas coefficient.
2. High Temperature Furnace – The highest temperature is 1000℃, and the stove heating rate can be adjustable from 1℃/min to 30℃/min.
3. Air Refrigeration – Open (shell type) furnace can be quickly cooling and optional refrigerator achieving rapid refrigeration.
4. Optional Cryotrap – The temperature can be rising from -100℃.
5. Temperature Control – Piping, steam generator, stove and TCD temperature can all be controlled by digit.
6. Thermal Conductivity Cell Detector – 4-arm rhenium tungsten TCD has extremely high linear, accuracy, sensitivity and stability.
7. Very small dead volume - Small volume valve and 3 mm pipe can reduce dead volume, thus reduce the deviation of the peak.
 
Easy Operation
1. Sample Support – U type pipe of large size can be convenient for sample loading and cleaning, and can add powder, rod or shot sample.
2. Sample Loading – Sample tube free of taking down is convenient to load samples. 
 
 
Safety
1. Heating Pipeline – All the pipeline, valve and liquid carburetor are heated to prevent condensation.
2. Safety – Independent furnace temperature thermal protection, sealing pressure reducing valve, current circuit breaker, fuse, the system design has timely security protection.
 
Parameter
  
Temperature System
 
Range:
70 to 1100 °C with CoolTec Cooler™
 
option, Ambient to 1100 °C (without CoolTec Cooler )
 
Selection:
Digitally set, 1 °C increments
 
Ramp Rates:
Up to 60 °C per minute within Ambient to 600 °C range
 
Up to 30 °C per minute from 600 to 800 °C range
 
Up to 10 °C per minute from 800 to 1100 °C range
 
 
 
Gases
 
Loop (Analysis):
H2, CO, O2, N2O, NH3,
 
vapors such as pyridine, water, etc.
 
Carrier:
He, Ar, and other gases
 
Preparation:
H2, O2, He, Ar, and others
 
 
 
 
Gas Flow Rate
 
All Mass Flow Controllers (MFCs)
 
Manual Control:
0 to 100 sccm
 
Automatic Analysis:
5 to 85 sccm Analysis:
 
 
 
Gas Delivery
 
Inlet Ports:
3 each for carrier gas,
 
treatment (analysis) gas
 
Temperature Control:
Internal gas lines and valves heated up to 150 °C
 
 
 
Sample Tube
 
Type:
Fused quartz FlowThru sample tubes,
 
for use up to 1100 °C, accepts powders and
 
pellets up to 8 mm in diameter
 
 
 
Electrical
 
Voltage:
85 to 265 VAC
 
Frequency:
50/60 Hz
 
Power:
900 VA, operating, max.
 
 
 
Environment
 
Temperature:
15 to 35 °C operating
 
0 to 50 °C non-operating
 
Humidity:
 
20 to 80% relative, non-condensing
 
 
 
 
 
 
 
 
 
 
 
 
 
Physical
 
Height:
65 cm
 
Width:
71 cm
 
Depth:
62 cm
 
Weight:
65 kg
 
 
 
 
Control Module Minimum Requirements
 
Processor:
Pentium 500 MHz or equivalent
 
Operating System:
Windows 2000 or XP Professional
 
Memory:
512 megabytes of RAM
 
Hard Disk Space:
2 gigabyte
 
Graphics Card:
1024 x 768
 
SVGA
 
Others
CD ROM drive & PCI interface 
 
Options
 
 
CoolTec Cooler
 
FID