Thermal Shock Chambers
Product Description:
The thermal shock series is a high-quality system proven with over 10 years of sales to major electronics, automotive, and aerospace manufacturers.
This series is distinguished by it's clean look, advanced touch-screen controller, and innovative hot/cold transfer system. Performance is able to meet strict Mil-Std test specifications, which require temperature recovery to be measured in the worst-case product sample.
With a temperature range of -65 to 150°C, the units are ready for the most extreme test applications. There are various size chambers available: 50L, 80L, 108L, 216L...
Technical parameters:
Thermal Shock Chambers
| Interal Dimension: | (W×H×D)50×50×60 cm |
| Structure: | Two chambers design(the hot chamber or the cold chamber) |
| Damper device: | Forced air drived damper |
| Chamber material: | SUS# Stainless steel plate |
| Refrigeration: | Cascade refrigeration system |
| Cooling method: | Water |
| Ambient temperature: | 0°~30° |
| Preheating temperature: | 60°C~+200°C |
| Precooling temperature: | -10°C~ -70°C |
| H.T. Shocking: | +60°C~+150°C |
| L.T. Shocking: | -10°C~-65°C |
| Temperature uniformity: | ±2°C |
| Simulated load ic | 5kg |
| HT IN-Zone | -65°C~+150°C/5min |
| LT IN-Zone | +150°C -65°C |
| Preheating time: | 45MIN |
| Precooling time: | 75MIN |
| Senor: | T×4 |
| Controller: | OYO-8226 |
| Setting range: | Temperature:-100°C~+200°C/TIME:OHIM~999H/CYCLE:0 ~9999 |
| Resolution: | Temperature:0.1°C |
| Output mode: | PID+SSR(control method) |
| Power supply | AC380V, 60Hz, three phase |
Mechannical / Refrigeration Cooling Method:
Thermal Shock Chambers
• Non-CFC cascade refrigeration system using semi-hermetic Tecumseh compressors
• Two ¾ hp blower motors circulate the conditioned air (one per zone)
• Cold storage mass maximizes cooling capacity.
• Standard dry air purge system extends the time between defrost cycles
• Hinged service access doors with lockablesnap-action latches
• Three levels of overheat protection and two levels of overcool protection
• Optional air-cooled condenser for sites where cooling water is not available.
• Two ¾ hp blower motors circulate the conditioned air (one per zone)
• Cold storage mass maximizes cooling capacity.
• Standard dry air purge system extends the time between defrost cycles
• Hinged service access doors with lockablesnap-action latches
• Three levels of overheat protection and two levels of overcool protection
• Optional air-cooled condenser for sites where cooling water is not available.
Product features:
Thermal Shock Chambers
- The machine is divided into three sections: the high temperature chamber, the low temperature chamber and the testing chamber.
- The color touch panel control system was made in Japan can display Chinese / English. The big size liquid-crystal LCD has been adopted.
- High formula memory capacity may be set to 120 patterns. The maximum cycle setting is 9999cycles. The maximum setting of each section is 999 Hrs 59 Mins.
- Provide with RS-232C communication interface device, capable of controlling / editing/ recording in the linkage with the computer and two set dynamic connection point and convenient application.
- Both 2 zone and 3 zone shocks can be chosen.
- All equipment is fully automatic, and all process are precisely controlled with P.L.C.’S.
- Of any part of the malfunctions, the machine will automatically shuts down and a warning light switches on.
- Machine can be shut down automatically in case of electrical power instabilities.