Steam Sterilization Autoclave SEKER 12L
Technical Paramater
1. Chamber Volume: 12Liter
2.Chamber Size: Ø 200× 360
3.Power: 1750W
4.Display: Digital Display
Features:
Autoclave 12L. It works according to the three times fractionated pre-vacuum method with a effective vacuum pump.
. High precision control system for more profect sterilization.
. Digital display.
. A built-in independent steam generator, making the sterilization fast and completely.
. Most of the main parts for TINGET autoclaves are imported from global famous company.
. Sterilization temperature: 121. C/1.1bar, 134. C/2.1bar.
. Great drying effect by vacuum.
. Open-type water storage tank which is very easy for filling and cleaning.
. Automatic water filling and draining.
Safety:
. Door protection locking system prevents the cycle from starting unless the door is properly locked, and also prevents the door from openning if there have pressuer in the chamber.
. Pressure safety valves prevent over pressurizeation in chamber and generator.
. A safety fuses prevent over heating in the chamber and generator.
. Able to detect and identify the exact cause of any problem by warning and showing a specific error code.
. Water level switch in the water tank, controlling the max and min water level.
. Have printer port, able to be connected TINGET printer to print the sterilization records for each cycle.
|
Program
|
Temperature(? C)
|
Pressure(bar)
|
Sterilziation Time(Min)
|
Total Time(Min)
|
|
SOLID
|
134
|
2.1
|
4
|
18-30
|
|
SOLID
|
121
|
1.1
|
20
|
30-45
|
|
WRAPPED
|
134
|
2.1
|
6
|
30-40
|
|
WRAPPED
|
121
|
1.1
|
20
|
35-50
|
|
COTTON
|
134
|
2.1
|
10
|
45-65
|
|
COTTON
|
121
|
1.1
|
30
|
50-75
|
|
PLASTIC
|
105
|
0.2
|
30
|
50-65
|
|
B& D TEST
|
134
|
2.1
|
3.5
|
22-35
|