Product Main

Specifications

 
Instructions 
Our portable PH meters display acid degree and temperature on the big LCD screen with 
excellent waterproof ,outdoor  applicability and ATC function.
The temperature sensor is directly immersed in the testing solution to ensure the fast and accurate numerical reading .
 
Use two point (PH 4.00,PH6.86) for auto calibration .
During calibration , put the PH meter in the standard solution and press "shift" button , Theinstrument will begin automatic calibration. 
Compared with the Analog circuit PH meter ,the Calibration speed and data accuracy are 
much more perfect..
It successfully avoid simulator adjust potentiometer calibration and repeatedly checking 
trouble.
The instruments' operation is simple, durable, economical and practical, completely suitable for' various industries acid-base testing needs.
Specifications 
1.Measuring range:           PH:  0.00-14.00 PH
                           T.:   0.1°C-80°C
2.Resolution:                PH:   0.01PH
                           T. :   0.1°C
3.Precision                  PH::+/- 0.01 PH
                           T. :  +/-0.5°C
4..Automatic temperature compensation:               0—80°C
5.Battery:                   3 *1.5 V(AG-13)
6.Operation temperature:       0---80°C
7..Calibration:               Automatic two  point calibration 
8.Auto shut off function:       4 minutes
9.Size :                     183mm*37mm*37mm
10.Weight :                  80g
 
 
Operation Steps
 
1.Take off the protective sleeve
 
2.Clean the Electrode with distilled water and wipe clean the spare water.
 
3..Press the "ON/OFF" button .Switch on the power.
 
4.Put the PH meter in the solution .Make sure the electrode  immersed in the solution.
 
5. Lightly shake the meter, read the data when the numeric stabilized.
 
  The temperate displays at the bottom of the screen .
 
6. Press the "ON/OFF" button to shut off  the instrument .
 
  Clean the electrode, cover the electrode protective sleeve.
 
Calibration 
 
 
Calibration 
1. Put proper 6.86, 4.00 and 9.18 standard solution apart in clean beakers 
2. In order to accurate calibration , put every buffer apart in two beakers.One for cleaning
    the electrode and the other for calibration. 
3. Press the switch,switch on the power supply.
4. Immerse the electrode in PH 6.86 standard buffer solution, gently shake the electrode 
    until the display digital stabilized.
5.Press long the "shift" button for 5 seconds, the instrument comes into the automatic 
   caliration Status (The instrument can recognize the PH 4.00 and PH 6.86 standard 
   solution until the display digital same as 6.86 ph of the standard solution. Then 
   complete the PH 6.86 point calibration
6.Clean the electrode with distilled water and dry it with filter paper.
7.Immerse the electrode in PH 4.00 standard buffer solution, gently shake the electrode 
   until the display digital stabilized..Press long "shift" button and repeat Step 5 .
   Complete the ph 4.00 point Calibration
8.Clean the electrode with distilled water and dry it with filter paper.Put the electrode in 
   the PH  9.18 standard solution and shake lightly until the data stable,The data is in 
   the range of allowable error.
9.This instrument calibration work is completed after all the steps.
Attention:
 
The instrument need recalibrate  in the following condition.
 
1.Long time no use or use very long time after calibration
 
2.The electrode is used Very frequently.
 
3. The testing accuracy requirement is very high.
 
4.Do not press the "shift" button long time .Otherwise the meter need recalibrate.
 
LowBatt 
 
When the display value is fuzzy or  unshown ,it shall check the battery voltage. Lessing
than 3 v,  it shall promptly replace the battery.
 
Unscrew  battery cover and replace the battery.  
 
Pay attention to the polarity of battery.
 
Warranty
 
The instrument  maintains for a year, from the delivery date , guaranteeing due to production and material defects 
 
If the instrument need to be changed or repaired ( not the artificial damage  or caused by 
fault using ) ,please mail the manual and the instrument together to our factory .
 
We will solve all the problem for you .