Hi-Pot Tester / Insulation Tester
Product Description
Features:
Compact size and light weight
Four test modes available, W, I, W-I, I-W
Linkable to Ground Bond Tester
Capable of storing 5sets of testing conditions
Electronic ramp up and dwell time for up to 999.9 seconds
Front panel lockout
Built-in PLC Remote interface
Front panel software calibration
Easy to operate
Ordering information:
7110 AC Withstand Voltage Tester
7112 AC Withstand Voltage / Insulation Resistance Tester
7120 AC/DC Withstand Voltage Tester
7122 AC/DC Withstand Voltage / Insulation Resistance Tester
Specification:
Product Description
Features:
Compact size and light weight
Four test modes available, W, I, W-I, I-W
Linkable to Ground Bond Tester
Capable of storing 5sets of testing conditions
Electronic ramp up and dwell time for up to 999.9 seconds
Front panel lockout
Built-in PLC Remote interface
Front panel software calibration
Easy to operate
Ordering information:
7110 AC Withstand Voltage Tester
7112 AC Withstand Voltage / Insulation Resistance Tester
7120 AC/DC Withstand Voltage Tester
7122 AC/DC Withstand Voltage / Insulation Resistance Tester
Specification:
| Model | LCX-7122 | LCX-7120 | LCX-7112 | LCX7110 | |||
| AC WITHSTAND VOLTAGE | |||||||
| Output Rating | 5KVAC/12mA | ||||||
| Range | Resolution | Accuracy | |||||
| Output Voltage, KVAC | 0-5.00 | 0.01 | ± (2% of setting + 5V) | ||||
| Output Frequency | 50Hz/60Hz ± 100ppm, User Selection | ||||||
| Output Waveform | Sine Wave, THD. < 2% (Resistive Load), Crest Factor = 1.3 - 1.5 | ||||||
| Output Regulation | ± (1% of output + 5V), From no load to full load | ||||||
| SETTINGS | |||||||
| Hi-Limit AC Current, mA | 0.10 - 12.00 | 0.01 | ± (2% of setting + 2counts) | ||||
| Lo-Limit AC Current, mA | 0.00 - 12.00 | 0.01 | ± (2% of setting + 2counts) | ||||
| Ramp Time, second | 0.1 - 999.9 | 0.1 | ± (0.1% + 0.05sec) | ||||
| Dwell Time, second | 0, 0.2 - 999.9 ( 0=continuous) | 0.1 | |||||
| Arc Detection | 0, 1 - 9 ranges (0=OFF, 9 is the most sensitivity) | ||||||
| DC WITHSTAND VOLTAGE (7120, 7122 ONLY) | |||||||
| Output Rating | 6KVDC/5mA | ||||||
| Output Voltage, KVDC | 0 - 6.00 | 0.01 | ± (2% of setting + 5V) | ||||
| Output Ripple | < 5%(6KV/5mA at Resistive Load) | ||||||
| SETTINGS | |||||||
| Hi-LimitDCCurrent, mA | 0.02 - 5.00 | 0.01 | ± (2% of setting + 2counts) | ||||
| Lo-LimitDCCurrent, mA | 0.00 - 5.00 | 0.01 | ± (2% of setting + 2counts) | ||||
| Ramp Time, second | 0.1 - 999.9 | 0.1 | ± (0.1% + 0.05sec) | ||||
| Dwell Time, second | 0, 0.2 - 999.9 (0=continuous) | 0.1 | |||||
| Arc Detection | 0, 1 - 9 ranges (0=OFF, 9 is the most sensitivity) | ||||||
| Discharge Time | < 200 msec | ||||||
| INSULATION RESISTANCE (7112, 7122 ONLY) | |||||||
| Output Rating | 1KVDC/1000MΩ | ||||||
| Output Voltage, KVDC | 0.10 - 1.00 | 0.01 | ± (2% of setting + 2V) | ||||
| SETTINGS | |||||||
| Hi-Limit Resistance, MΩ | 0, 1 - 1000 (0=OFF) | 1 | ± (3% of setting+2counts)at > 500VDC ± (7% of setting +2counts)at < 500VDC | ||||
| Lo-Limit Resistance, MΩ | 1 - 1000 | 1 | |||||
| Delay Time, second | 0, 0.5 - 999.9 (0=continuous) | 0.1 | ± (0.1% + 0.05sec) | ||||
| MEASUREMENT | |||||||
| AC/DC Voltage, KV | 0 - 6.00 | 0.01 | ± (2% of reading + 1count) | ||||
| DC Voltage, KV (IR only) | 0.10 - 1.00 | 0.01 | ± (2% of reading + 2V) | ||||
| AC Current, mA | 0.10 - 12.00 | 0.01 | ± (2% of reading + 2counts) | ||||
| DC Current, mA | 0.02 - 5.00 | 0.01 | ± (2% of reading + 2counts) | ||||
| Resistance, , MΩ | 1 - 1000 (Auto Range) | 0.001 0.01 0.1 | ± (3% of reading+2counts)at > 500VDC ± (7% of reading +2counts)at < 500VDC | ||||
| GENERAL | |||||||
| Input Voltage AC | 110/220Vac± 15%, 50/60Hz ± 5%, Fuse 3A Fast | ||||||
| PLC Remote Control | Input: Test, Reset | ||||||
| Output: Pass, Fail, Processing | |||||||
| Memory | 5 Sets Store/Recall of each Memory, 4 Test Mode Selectable (W, I, W∼ I, I∼ W Only 7112, 7122) | ||||||
| Display | 16× 2 LCD with back light | ||||||
| Key Lock | To prevent unauthorized alteration of the test Parameters | ||||||
| Calibration | Build-in software and external calibrated meters | ||||||
| Environment | 0-40º C, 20-80%RH | ||||||
| Dimension/Net Weight | 280mm(W)× 89mm(H)× 370mm(D)/ 9Kg | ||||||
| STANDARD ACCESSORIES | |||||||
| Power Cord(10A) | × 1 | ||||||
| Fuses | × 3 (Including a spare contained in the fuse holder) | ||||||
| High Voltage Test Cable(1101) | × 1 | ||||||
| Return Test Cable(1102) | × 1 | ||||||