1. Function
Charpy Impact Test Low Temperature Chamber is a supplemental equipment for impact test under low temperature.
2. Features
1) It adopts imported overlap refrigeration technology, heat balance principle and loop mixing method;
2) It can achieve automatic uniform cooling and constant temperature for specimen;
3)It complies with the temperature control provisions of relevant national standards and GB/T 229-2007 “Metal Charpy Notch Impact Test Method”.
4) It can meet all requirements for specimen testing at low or ultra-low temperature with easy operation, reliable quality and high working efficiency. Besides, it can also be used for other low-temperature detection and testing.
5) There are also CDW-40, CDW-60, CDW-80, CDW-100, CDW-150, and CDW-19(liquid nitrogen) models to meet needs from different customers.
3. Main technical parameters
3. Main technical parameters
Model
|
CDW-80T Impact Test Low Temperature Chamber
|
|
Cooling range
|
+30~-80°C (room temperature:≤25°C)
|
|
Constant temperature accuracy
|
±0.5°C
|
|
Cooling speed
|
+30°C~0°C
|
About 2.0°C/min
|
0°c~-20°C
|
About 1.5°C/min
|
|
-20°C~-60°C
|
About 1.0°C/min
|
|
|
-60°C~-80°C
|
About 0.7°C/min
|
Max. Dimension
|
900×505×850 mm
|
|
Available working place
|
150×150×240mm
|
|
Specimens quantity
|
60<quantity<120(Impact Specimens Size: 10×10×55mm)
|
|
Digital timer
|
1min~99min(Resolution:1min)
|
|
Cooling medium
|
alcohol or other unfrozen liquid
|
|
Stirring Motor
|
12w
|
|
Power supply
|
220V~240V, 50HZ, 7KW
|
|
Work temperature
|
≤25°C
|