Basic Part
Mainly parameter:
With single table and single tube with radiographic function.
Max mA: 200 mA Max voltage: 125KV
Warranty: 12 months after shipment date.
Delivery time: within 4 weeks after payment.
Package: export wooden
Standard configuration:
1 controller
1 generator
1 radiographic table
1 tube
1 floor stand
2 high-voltage cables
Features:
1. Single diagnostic table with X-ray tube and single radiographic table with X-ray tube configuration.
2. Single phased bridge-type full wave and high voltage commutation
3. In order for the X-Ray tube to be used safely for a long period of time, multi-layer protection are utilized. These include: X-Ray tube capacity protection, revolving anode-starting protection, radiographic timer-circuit protection, fluoroscopy time accumulation device protection and X-Ray tube overheating protection.
4. Various specifications of spot films can be used, with high level ease of operation. Radiation protection and diffusion protection devices are utilized to ensure operators’ and examinee’ safety.
5. CDGI multi-purpose radiographic table has trackless twin rope collapse protection floor stand system. The tabletop allows both vertical and horizontal movement and the X-Ray tube assembly can also move and revolve with ease. It is equipped with an electromagnetic braking system and allows the use of cassettes with varying specifications.
6. Outlet for tomographic devices.
7. The design is intended for small to medium hospitals to be used in conjunction with X-Ray fluoroscopy, spot film radiography, bucky radiography, tomography and chest film radiography.
Power:
A.C. 380V(220V)±10%, frequency 50HZ±1HZ, Capacity≥20KW; Power resistance:380V≤1Ω, 220V≤0.34Ω,Earth resistance≤4Ω
High Voltage Generator:
Radiography: 25-200mA divided adjustable, 50-100(125)KV, step enters: 1 KV, 0.04-6.3s, totally 23steps are adjustable
X-Ray Tube Assembly:
Small focus: 1×1, 20KW, Large focus:2×2, 40KW, normal revolving speed 2800r/min, heat capacity 140KHU
Collimator:
Electro-motion
Multi-leaf’s type Multi-function Radiographic Table:
Movement of the tabletop: Longitudinal ±450mm, transverse: ±100mm
The tabletop height: 700mm
Moving grid: Grid density: 28L/cm, Grid ration: 8:1, Focusing distance: 100mm, moving range: 640mm
Film’s specifications:5″×7″,8″×10″,10″×12″,11″×14″,12″×15″,14″×14″,14″×17″
Movement of the floor tube stand: ±180°, Location/90°, Longitudinal 1500mm
Movement of X-Ray tube assembly: To go around the horizontal axle ±180°, up and down: 1100mm
Floor tube stand:
Moving: up and down 300-1800mm, longitudinal: 3000mm, transverse: 780mm
X-Ray tube assembly rotary: to go around the floor tube stand: ±180°location per 90°, to revolve ±180°round the horizontal axles and revolve 120°round the vertical axles itself
Floor tube stand height: 2750-3150mm