Quick Details
Flow: 0.001 – 9.999 ml/min
Max Pressure: 10000 psi
Number of Solvent: 1
Minimum detection limit: 1×10-9g/ml
Brand Name: Surwit
Model: LC-10Tvp
Certificate:GB/T 19001-2008/ISO 9001:2008
Series Pump
GLP/GMP, FDA, 21CFR Part Ⅱ standard.
Features and Specification
LC-10Tvp Liquid Pump
Infusion way
|
Series dual-plunger pump
|
Pressure Accuracy
|
± 2%
|
Flow Rates
|
0.001 ~9.999 ml/min
|
Flow Accuracy
|
≤0.5%
|
Pressure
|
0 - 9999 psi
|
Flow Precision
|
≤0.2% RSD
|
Pulsation
|
<15psi from 600 to 1,600 psi, Flow 1mL/min
|
Operating Temperature
|
4~40℃
|
Dimensions
|
130H x 260W x 420 D, (mm)
|
Weight.
|
11 kg
|
SPD-10Tvp UV-VIS Spectrophotometric Detector
Wavelength
|
190-700 nm
|
Spectral bandwidth
|
6 nm
|
Wavelength Accuracy
|
≤1 nm
|
Flow pool volume
|
8 μ L
|
Wavelength Reproducibility
|
≤± 0.1 nm
|
Optical path
|
10 mm
|
Dynamic Noise
|
≤± 0.75×10-5AU (Methanol, 1ml/min, 254nm, 20°C)
|
Static Noise
|
≤± 0.5×10-5AU(Empty pool, Response time 1 s, 20 °C)
|
Dynamic Drifting
|
≤± 1×10-4AU (Methanol, 1ml/min, 254nm, 20°C)
|
Static Drifting
|
≤± 0.5×10-4AU(Empty pool, Response time 1 s, 20 °C)
|
Qualitative repeatability
|
RSD6 ≤ 0.1%
|
Quantitative repeatability
|
RSD6 ≤ 0.5%
|
Dimensions
|
130H x 260W x 420 D, (mm)
|
Weight.
|
(11 kg)
|
Operating Temperature
|
4~40°C
|
|
|
LC-10Tvp Isocratic Standard Configuration
LC-10Tvp Liquid Micro volume series double plunger
|
1 set
|
SPD-10Tvp UV-VIS Spectrophotometric Detector
|
1 set
|
SCL-10Tvp System Controller
|
1 set
|
7725i manual injector
|
1 set
|
Chromatography data system
|
1 set
|
Chromatography column
|
1 set
|
Packaging & Delivery
Packaging Details: TNT, UPS, FedEx available. Or per your requirements.
Delivery Detail: 7 days by air
Payment: PREPAID 100% in advance
Item Description
The LC-10Tvp Liquid Chromatograph System is designed by the Research and development team of Surwit Technology Inc.. It can be widely used in the field of medicine, food, chemical industry and environment.
In either case, the system incorporates several key features which make it truly one of a kind in the world of HPLC Systems. The LC-10 liquid system is more durable and stable as all the key parts are imported from USA.
Trinity design
LC-10 Tvp adopted the trinity system of column over, system controller and solvent tray as a whole, which can reasonably used the space and enhanced the management of temperature and solvent.
Audit trail and e-signature
Conform to GLP/GMP, FDA, 21CFR Part Ⅱ standard.
Intuitive Software Control
The LC-10Tvp Liquid Chromatograph System adopts the most advanced liquid chromatography system called VI2010 HPLC Data WorkStation, which designed for international standard, all the action of LC-10T can be directly controlled by the computer. The control is designed with the dedicated keys that allow even a novice to turn on the pump and create methods. An easy to read data display allows the user to monitor several parameters from the same screen. When configured as a gradient HPLC system user can program and store different methods, all the methods can be linked or run independently.
Easy maintenance
All user serviceable components are easily accessed from the front panel. Once the head assembly is removed, the pistons and seals can be replaced quickly and easily without the use of tools. The highest voltage of LC-10Tvp can be reached to 10000 psi, and the minimum detection limit is 1×10-9g/ml.