Product Main

Quick Details

Condition:
Place of Origin:
Brand Name:
Model Number:
Voltage:
Power(W):
Dimension(L*W*H):
Weight:
Certification:
Warranty:
After-sales Service Provided:

Specifications

LDS series Multi-effect water distiller adopted the purified water as the feed water , produced WFI by heating industrial steam. The product satisfies the requirement of CGMP(CP2010), USP34AND EP7.

Working principle
The qualified feed water is input into the condenser after increased the pressure by multilevel pump. And then enters into each preheater(the feed water will be heated by the secondary steam and distilled water in the condenser, in the first stage preheater, the feed water heated by the industrial steam condensate water, after enters the other stages preheater, it is heated by the secondary steam condensate water). Next the feed water in input into the water distributor of the 1st stage evaporator, which is spouted to the inside of the heating tubes and flowed as film. The produced second steam with drips enters into the gas and steam separating setting from the bottom of the heating tubes. The separated pure steam enters into the next stage as heating steam, meanwhile, the feed water, which is not evaporated, enters into the next stage. And then it is repeated above mentioned process. The pure steam from last stage is converged with each stage condensed water(except 1" stage), after cooling and drainage of the non-condensable gas, (according to different model, there should be a device for discharge the non-condensable gas continuously) it is became the WFI. There is an on-line conductivity test at the output orifice, the qualified distilled water will be output as the WFI, and the unqualified will be discharged.

Water- output(L/h)
 
 
 
Multi-effect
Raw steam consumption
L/h
Pure steam output
L/h
 
Distilled water outpet
L/h
Raw water
consumption
L/h
Cooling water
consumption
L/h
Power
KW
50
3
121
34
59
65
0
0.37
300
3
155
135
>350
420
0
0.75
4
120
0
500
4
195
210
>600
720
0
1.1
5
165
0
6
140
0
1000
4
390
320
>1200
1500
0
1.5
5
420
0
6
295
0
1500
4
590
425
>1800
2160
0
1.5
5
500
0
6
450
0
2000
4
790
525
>2500
3000
0
2.2
5
670
0
6
600
0
3000
4
1180
640
>3600
4320
0
 
5
1000
0
6
900
0
Raw steam consumption pressure
(
MPa )
(work pressure 0.2~0.5MPa, design pressure 0.55MPa)
Raw&Cooling water quality
deionized water, electric conductivity<5μS/cm
Raw&Cooling water temperature
15~25℃
Distilled water temperature
92~98℃